5.4.4 Activating the Channels Output

5.4.5 Store and Recall

User Manual 5601.1343.02 ─ 02
A delay can be applied to every channel fuse. Also, individual electronic fuses (Fuse-
Link) can be logically linked so as to switch off the interlinked channels.
For more information, see
The outputs of all the channels (Ch 1, Ch 2, Ch 3) can be switched on or off by tog-
gling the Output key on the front panel.
By default, the output is switched off when the instrument is switched on. The output is
also automatically switched off when no channel is selected. This design prevents a
connected load from being damaged unintentionally.
1. Press the desired channels.
Selected channel keys (Ch 1, Ch 2, Ch 3) illuminate.
2. Press Output key.
The R&S NGE100B output the set voltage of the selected channel.
Depending on the output mode, the font text in the channel display area shows
green when in CV mode and red when in CC mode.
See
Chapter 5.3, "Output
Figure 5-12: Output of Ch 3 in CV and CC mode
Multiple outputs can be switched on at the same time. See more information in
ter 7.1, "Parallel and Serial
See also
Power
Derating.
Upon power-up, the instrument loads the stored setting from memory location M1 with
the exception on EasyArb parameters. Auto saved parameters are also applied during
startup.
The R&S NGE100B output state of all channels (Ch 1, Ch 2, Ch 3) are disabled when
the recall function is activated.
Fuse
for the fuse delay and FuseLink configuration.
Modes",
on page 34.
Mode",
on page 98.
Allice Messtechnik GmbH
Instrument Functions
Basic Functions
Chap-
38