11. Set the "Rep" to configure repetition cycle for the arbitrary data.
By default, the repetition cycle is set to infinity.
12. Set the "End Behavior" to handle the way to end the automation of the arbitrary
function.
13. Select "Save" (new file) or "Save as" (existing file) to save the arbitrary data.
The R&S NGL/NGM displays a popup message to shows that file saved success-
fully.

6.7.2 Ramp

The "Ramp" function configures a constant rise of supply voltage within a set time
frame. The output voltage can be increased continuously within a 10 ms to 10 s time
frame. Each channel has an independent ramp configurations. See
Figure 6-24: Ramping voltage output
1. Press [Settings] key.
The R&S NGL/NGM displays the device/channel menu window.
2. Select the required channel tab to configure ramp function.
The R&S NGL/NGM displays the selected channel menu.
3. Select "Ramp" from the menu.
User Manual 1178.8736.02 ─ 05
"Off": The output of the selected channel is turned off after performing the arbi-
trary function.
"Hold": The last voltage and current values remains at the output terminal of
the instrument.
Instrument Functions
Advanced Features
Figure
6-24.
66