Activating The Channels Output; Store And Recall - R&S NGE100B Series User Manual

Hide thumbs Also See for NGE100B Series:
Table of Contents

Advertisement

®
R&S
NGE100B
A delay can be applied to every channel fuse. Also, individual electronic fuses (Fuse-
Link) can be logically linked so as to switch off the interlinked channels.
For more information, see

5.4.4 Activating the Channels Output

The outputs of all the channels (Ch 1, Ch 2, Ch 3) can be switched on or off by tog-
gling the Output key on the front panel.
By default, the output is switched off when the instrument is switched on. The output is
also automatically switched off when no channel is selected. This design prevents a
connected load from being damaged unintentionally.
1. Press the desired channels.
Selected channel keys (Ch 1, Ch 2, Ch 3) illuminate.
2. Press Output key.
The R&S NGE100B output the set voltage of the selected channel.
Depending on the output mode, the font text in the channel display area shows
green when in CV mode and red when in CC mode.
See
Figure 5-12: Output of Ch 3 in CV and CC mode
Multiple outputs can be switched on at the same time. See more information in
ter 7.1, "Parallel and Serial
See also

5.4.5 Store and Recall

Upon power-up, the instrument loads the stored setting from memory location M1 with
the exception on EasyArb parameters. Auto saved parameters are also applied during
startup.
The R&S NGE100B output state of all channels (Ch 1, Ch 2, Ch 3) are disabled when
the recall function is activated.
User Manual 5601.1343.02 ─ 04
Fuse
Chapter 5.3, "Output
Modes",
Mode",
Power
Derating.
for the fuse delay and FuseLink configuration.
on page 35.
on page 98.
Instrument Functions
Basic Functions
Chap-
39

Advertisement

Table of Contents
loading

This manual is also suitable for:

Nge103bNge102b

Table of Contents