R&S
NGL200/NGM200
Figure 6-16: Over power protection dialog
2. Activate the "Enabled" menu item.
The R&S NGL/NGM enables the OPP and displays the "Over Power Protection
(OPP)" icon on the selected channel status bar information.
3. Set the required level for OPP.
The R&S NGL/NGM displays the on-screen keypad to set the value.
4. Confirm value with the unit key (mW or W).

6.5.4 Safety Limits

With safety limits set in the instrument, the range of the output voltage and/or output
current can be limited. The safety limit prevents inadvertently setting values dangerous
for the connected DUT.
1. Select "Safety Limits" menu item from the menu.
The R&S NGL/NGM displays the "Safety Limits" dialog.
Figure 6-17: Safety limits dialog
User Manual 1178.8736.02 ─ 05
Instrument Functions
Protection
57