R&S NGL200 Series User Manual page 130

Hide thumbs Also See for NGL200 Series:
Table of Contents

Advertisement

®
R&S
NGL200/NGM200
Example:
[SOURce:]POWer:PROTection:CLEar
Resets the OPP state of the selected channel. If an OPP event has occurred before,
the reset also erases the message on the display.
Example:
Usage:
[SOURce:]POWer:PROTection:LEVel <power>
[SOURce:]POWer:PROTection:LEVel?
Sets or queries the overvoltage protection value of the selected channel.
Parameters:
<power>
[SOURce:]POWer:PROTection:TRIPped?
Queries the OPP state of the selected channel.
Example:
Usage:
User Manual 1178.8736.02 ─ 05
OFF | 0
OPP is deactivated
ON | 1
OPP is activated
See
Example "Configuring the overpower protection"
on page 126.
See
Example "Configuring the overpower protection"
on page 126.
Event
numeric | MIN | MAX | DEFault
numeric
Numeric value of the power protection level in watts.
MINimum
Minimum value of the power protection level at 0.00 W.
MAXimum
Maximum value of the power protection level at 60.00 W.
Range:
0.00 to 60.00
*RST:
60.00
Default unit: W
See
Example "Configuring the overvoltage protection"
on page 123.
POW:PROT:TRIP?
Response 1, the OPP is tripped.
Response 0, the OPP is not tripped.
Query only
Remote Control Commands
Configuration Commands
127