Current Leakage Check Failed

Warning Message: "Current leakage check failed."
Possible Causes
Dirty electrode block.

Chip Temperature Warning

Warning Message: "Average chip temperature is out of range."
Possible Causes
Chip heater failed.
V4.2
Recommended Actions
Clean the O-rings (see
Run the Current Leak Diagnostic Test (see
If the problem is not resolved, contact PerkinElmer
Technical Support (see
Recommended Actions
Restart the instrument. If problem persists, contact
PerkinElmer Technical Support (see
LabChip GX User Manual
Troubleshooting and Diagnostics 289
page
281).
page
3).
page
page
291).
3).