Current Leakage Check Failed
Warning Message: "Current leakage check failed."
Possible Causes
Dirty electrode block.
Chip Temperature Warning
Warning Message: "Average chip temperature is out of range."
Possible Causes
Chip heater failed.
V4.2
Recommended Actions
Clean the O-rings (see
Run the Current Leak Diagnostic Test (see
If the problem is not resolved, contact PerkinElmer
Technical Support (see
Recommended Actions
Restart the instrument. If problem persists, contact
PerkinElmer Technical Support (see