Verify Positions Window

Use the Verify Positions Window to verify the taught reference
position and the settings for the plate selected in the
Window.
The Verify Positions window contains the following options and
buttons:
Option/Button
Verify Plate
Position
Verify Buffer
Position
Verify Ladder
Position
Test button
Back button
Next button
Cancel button
V4.2
Figure 106. Verify Positions Window
Function
If selected, the plate holder moves so that the sipper pierces
the tape in the well each corner of the plate. If the well is not
pierced in the center of each well, click the Back button and
adjust the Plate Information.
If selected, the plate holder moves so that the sipper pierces
the tape over the buffer vial. If the buffer vial is not pierced in
the center, verify and adjust the Reference Position.
If selected, the plate holder moves so that the sipper pierces
the tape over the ladder vial. If the ladder vial is not pierced in
the center, verify and adjust the Reference Position.
Moves the plate holder to test the selected positions.
Displays the
Verify a Plate
Displays the
Finished
Closes the Robot Teach Wizard without saving changes to any
positions.
LabChip GX User Manual
Software Reference 239
Window.
Window.
Verify a Plate