Specifications
This section lists the technical specifications for the LabChip GX
instruments. Technical specifications are subject to change without
notice.
General
Size (H x W x D)
Weight
Plate Capacity
Ventilation/Cooling 3" (77 mm) minimum space required around instrument for
Environmental
Operating Temperature
Operating Humidity
Storage Temperature
Transient Overvoltages
Pollution
Altitude
Indoor Use Only
Electrical
Power Input
Line Voltage
Total Heat Emission
Fuses
Grounding
Computer Interface
V4.2
18.75 in. (48cm) x 19.25 in. (49cm) x 25.375in.(64cm)
88 lbs. (40 kg)
one 96-well or 384-well microplate
proper air flow.
65° to 78°F (18° to 26°C)
20% to 70% relative humidity, noncondensing
50° to 104°F (10° to 40°C)
Installation Category II Overvoltage
Pollution Degree 2
Up to 2000m
3.0A max at 100-127Vac, 50-60Hz (±10%)
1.5A max at 200-240Vac, 50-60Hz (±10%)
100-127/200-240V~ nominal
90-264V~ maximum operating range
800W
No customer-replaceable fuses in system. Contact
PerkinElmer Technical Support (see
are suspected
Through the power cord
USB 2.0
LabChip GX User Manual