Analytical Pump (100 Μl Piston) Flow Check @ 1Ml/Min - PerkinElmer FX-10 UHPLC Manual

Liquid chromatograph flexar analytical
Table of Contents

Advertisement

It is important that the volume delivered is correct; the pressure during the delivery is not
critical. There is a minimum pressure requirement of 350 psi for the check valves to
operate properly.
Test Steps
Connect the backpressure regulator to the output of the pump. This device will create the
necessary back pressure on the pump check valves.
Purge the pump and fill the solvent line to be used for the test with water.
Zero the scale. Weigh the collection vessel on a calibrated scale or balance.
Take the collection vessel and go to the pump and set the pump flow for 1 mL/minute.
Start the pump. (Direct the initial flow of water off to waste until the pressure stabilizes.
When pressure stabilizes, direct the flow of water into the collection vessel and
simultaneously start a stopwatch or other timing device.
Let the pump run for 5 minutes.
When 5 minutes has elapsed, remove the tube from above the collection vessel and
insert it into the waste.
Weigh the collection vessel with the water. Subtract the weight of the collection vessel.
The remaining weight of the water should be 5 grams ± 0 .05 g (40.95 to 5.05 g)
1 mL/min * 5 min = 5 g (water weighs 1 gram per mL)
Divide the weight of the water by 5 to get the volume delivered in a 1 minute interval.
Record this number in the Test Result field of Flow Accuracy @ 1 mL/min portion of the
IQOQ Test Certificate. Enter N/A if testing FX-10 UHPLC Pump.
2.4.2
Test Results
Parameter
Analytical Pump (100 µL
Piston) Accuracy: 1
mL/min
2.4.3 Analytical Pump (100 µ µ µ µ L Piston) Flow Check @ 3 mL/min
Test Conditions
Chromatographic Conditions:
Test Prerequisites
It is important that the volume delivered is correct; the pressure during the delivery is not
critical. There is a minimum pressure requirement of 350 psi for the check valves to
operate properly.
Comments:
PerkinElmer LAS Representative:
Reviewed By:
Flexar Analytical / FX-10 UHPLC Pump IQ/OQ Rev. D June 2009
Specification
± 1% of 1mL/min
(00.99ml/min to
1.01ml/min)
100 % H 2 O
3 mL/min
Test Result
Pass/Fail
Date:
Date: