Teach the X, Y and Z Reference Positions Window

Use the Teach X, Y, and Z Reference Positions Window to align the
sipper at the reference position, save the reference position, and
test the alignment.
Figure 104. Teach the X, Y and Z Reference Positions Window
The Teach the X, Y, and Z Reference Positions Window contains
the following options and buttons:
Option/Button
Left/Right buttons
Forward/Backward
buttons
Up/Down buttons
Step Size options
Set button
Test button
V4.2
Function
Moves the plate holder toward the left or right side of the
instrument. The distance the plate holder moves is set in the
Step Size option below the buttons.
Moves the plate holder toward the front or back of the
instrument. The distance the plate holder moves is set in the
Step Size option below the buttons.
Moves the plate holder up toward the top of the instrument or
down toward the bottom of the instrument.
Select the desired distance to move the plate holder each time
the positioning buttons are clicked. Use the Step size buttons
under the positioning buttons for the axis that is being moved.
Sets the current position of the plate holder as the reference
position.
Moves the plate holder to verify that the reference position is
set properly. If set properly, the sipper moves into the hole on
the plate holder without hitting the sides of the hole.
LabChip GX User Manual
Software Reference 237