Electrical Safety; Power Cord Selection - PerkinElmer LabChip GX User Manual

Table of Contents

Advertisement

Electrical Safety

The LabChip GX is powered by a UL/CSA/VDE approved 100-240
VAC, 50/60 Hz input, 5, 15, 24 VDC output power supply.
Additionally, the LabChip GX High Voltage circuitry is current-
limited to non-hazardous levels. Users should observe the
following:
WARNING
Do not open the instrument enclosure. There are no user
serviceable parts inside.
The wall outlet or the power cable connector on the back of the
instrument should be accessible after the system's installation, to
enable trained service personnel to safely disconnect power from
the system during servicing.
The computer supplied with the LabChip GX instrument has internal
lithium batteries. Batteries should not be incinerated.
WARNING
Danger of explosion if battery is incorrectly replaced. Replace only
with the same or equivalent type recommended by the
manufacturer's instructions.

Power Cord Selection

United States and Canada
The LabChip GX instrument is shipped with a NEMA 5-15 / IEC 320
power cord. If the power cord needs to be replaced, substitute
power cords must be UL Listed, Type SJT or equivalent, minimum
No. 18 AWG, 3-conductor with ground conductor that for safety
considerations should never be disconnected or defeated. The
cord's plug to the wall must be a three-pin grounding type connector
with a NEMA 5-15P (15A, 125V) plug configuration. The cord's
connector at the unit must conform to requirements for an EN 60
320/IEC 320 Standard Sheet C13 connector.
The equipment is intended to be plugged into a standard NEMA 5-
15R receptacle in the wall.
V4.2
LabChip GX User Manual
Instrument Safety
10
PerkinElmer