Move Timeout
Error Message: "Robot Failure."
Possible Causes
1. Robot failure
Pressure Leak Detected
Error Message: "Pressure leak detected. Check and clean the O-
rings at the chip interface."
Possible Causes
1. O-Ring missing or
dirty.
Focus Failed
Error Messages:
•
•
•
•
Possible Causes
1. Focus nominal starting
position is incorrect.
2. Laser or detector failure.
3. Laser interlock switch
failure.
4. Bad or jammed focus
motor.
V4.2
Recommended Actions
Check for an object blocking the robot motion and
remove the blockage.
Restart the instrument.
If the problem is not resolved, contact PerkinElmer
Technical Support (see
Recommended Actions
Clean the O-rings (see
Run the Pressure Diagnostic Tests (see
If the problem is not resolved, contact PerkinElmer
Technical Support (see
"Focus failed due to less than 2 peaks found during horizontal
scan."
"Focus failed due to more than 3 peaks found during horizontal
scan."
"Focus failed due to dry focus marks not found."
"Focus failed due to error in horizontal scan."
Recommended Actions
Run the Optics Diagnostics tests (see
If the problem is not resolved, contact PerkinElmer
Technical Support (see
LabChip GX User Manual
Troubleshooting and Diagnostics 287
page
3).
page
281).
page
page
3).
page
page
3).
291).
291).
PerkinElmer