Print Validation Reports Window - PerkinElmer LabChip GX User Manual

Table of Contents

Advertisement

Print Validation Reports Window

Use the Print Validation Reports Window to print the results after
performing Installation Qualification (IQ) or Operational
Qualification (OQ). OQ is performed in the
Window.
To open the Print Validation Reports Window, select Validation 
Reports on the
The Print Validation Reports Window contains:
Option/Button
Diagnostics Report
Software IQ Report
Failed Tests Only
Report Period
Print Preview
button
Print button
Close button
V4.2
LabChip GX Main
Figure 100. Print Validation Reports Window
Function
If selected, the results of Hardware OQ tests are printed.
If selected, the results of Software IQ tests are printed.
If selected, only tests that have failed are printed. If not
selected, both passed and failed tests are printed.
Specifies the dates when the tests were run. Tests performed
between the From and To dates (inclusive) are printed.
Displays a preview of the report and enables you to print or
export the results.
Opens the Print window to print the report.
Closes the window.
LabChip GX User Manual
Software Reference 232
System Diagnostics
Window.
PerkinElmer

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Labchip gx ii

Table of Contents