Monitoring the Run

The following occurs after a run is started in the LabChip GX
software:
1
2
3
4
5
6
7
To save the analyzed data:
1
2
To view or re-analyze the data, see
V4.2
The priming and warming steps are performed.
The priming step fills the channels of the chip with reagent.
(Only performed if the chip holder has been opened since
the last run.)
The warming step allows the heater plate located in the chip
holder to regulate chip temperature to 30°C.
Data collection begins after the warming and priming steps
are completed.
After the run begins, the Start button on the LabChip GX main
window changes to Stop.
The
Active Data Tab
currently being read. Data is saved to a file with the name
shown above the plate diagram after each well is completed.
To view the results for individual wells as data is acquired or
after the run is finished, click a well in the
List, a sample name in the
View. Data from the selected well displays in the
Electropherograms Tab
For more information on data analysis, see
page
42.
To stop the run before it is complete, see
When the assay is complete, Run Successful displays in the
Status line.
If desired, remove the plate and/or remove the chip.
Select File  Save Workspace. The
Window
opens with the default workspace name as the name of
the data file.
If desired, change the location and/or the name of the
workspace file and click the Save button. (Workspace files have
a .gxw file extension.)
LabChip GX User Manual
displays the electropherogram of the well
Well Table
View, or a lane in the
and the
Peak Table
Save Workspace As
Data
Analysis.
Operation
Plate View or Plate
Gel
Overlay
View.
"Data Analysis" on
Stopping a
Run.
PerkinElmer
30