Titer Tab

Use the Titer tab on the
in the plate that contains the standard to be used as the calibration
standard. The
(Titer is not available for Glycan or Protein Charge Variant assays.)
The Titer tab contains the following options and buttons:
Titer Table - Lists the standard size, well locations, and
concentrations for the titer on the plate.
V4.2
Titer
is used for calibration instead of the ladder.
Figure 80. Assay Analysis Window - Titer Tab
Size - Specifies the size of the standard in BP for DNA, kb for
gDNA, nt for RNA, or kDa for Protein assays.
Window (%) - Specifies the tolerance window as a percent of
the size for the standard to allow for small variations in size.
Found in Well- Specifies the location of the standard in the
plate. Type the well number or double-click to select the well in
the plate diagram. (Not displayed for imported titers.)
Normalized Area - Specifies the titer peak normalized area for
imported titers. The table for imported titers is Read Only and
cannot be edited. (Not displayed for manually entered titers.)
Concentration - Specifies the concentration of the standard in
the specified well.
LabChip GX User Manual
Assay Analysis Window
Software Reference 202
to specify the wells