Lam bda 365 5-Pos Variable P athlength Cell Holder
This instruction sheet describes the installation of this accessory which is used with the Lambda 365
Spectrophotometer.
Read these instructions before you install this accessory.
NOTE:
Contacting P erkinElm er
Supplies, replacement parts, and accessories can be ordered directly from PerkinElmer, using the part
numbers.
See our website:
http://perkinelmer.com
PerkinElmer's catalog service offers a full selection of high-quality supplies.
To place an order for supplies and many replacement parts, request a free catalog, or ask for information:
If you are located within the U.S., call toll free 1-800-762-4002, 8 a.m. to 8 p.m. EST. Your order will be
shipped promptly, usually within 24 hours.
If you are located outside of the U.S., call your local PerkinElmer sales or service office.
Features
• Available with various pathlength cells (5, 10, 20, 40, 50 mm)
• Excellent durability
Figure 1 Lambda 365 5-Pos Variable Pathlength Cell Holder [P/N: 4100104]
I nstallation I nstructions
PerkinElmer, 710 Bridgeport Avenue,
Shelton, CT06484-4794, U.S.A
Produced in the USA.
09931462A

Advertisement

loading

Summary of Contents for PerkinElmer Lambda 365

  • Page 1 09931462A Lam bda 365 5-Pos Variable P athlength Cell Holder I nstallation I nstructions This instruction sheet describes the installation of this accessory which is used with the Lambda 365 Spectrophotometer. Read these instructions before you install this accessory. NOTE:...
  • Page 2: Specification

    09931462A Dim ensions and Specifications Dim ensions Physical Characteristic Specification Comment Height (mm) 139.5 Outline Width (mm) 134.5 Depth (mm) Height (mm) Suitable for the Standard Inner Width (mm) Cells Depth (mm) 12.6 Weight (kg) Specification Physical Characteristic Specification Comment Space between cells (mm) Moving distance of one cell...
  • Page 3 09931462A Configuration of the 5-P os Variable P athlength Cell Holder Figure 2 Lambda 365 5-Pos Variable Pathlength Cell Holder Variable Partition Variable Pathlength Cuvette Holder ① 5 mm cell ② 10 mm cell ③ 20 mm cell ④ 40 mm cell ⑤...
  • Page 4: Installation

    09931462A - Used to fix a cell holder, a base plate or a front plate for Lambda 365. - Spare screws (2ea) are enveloped with the accessory. Figure 4 Phillips round head screw with washer (M4 *12L) I nstallation Make sure the instrument is turned off while installing this accessory.
  • Page 5 09931462A 4. Pull out the cell holder and base plate by hand. Figure 7 Pulling out the cell holder and base plate 5. Insert the 5-Pos Variable Pathlength Cell Holder in the sample compartment. Figure 8 Inserting the 5-Pos Variable Pathlength Cell Holder in the sample compartment...
  • Page 6 09931462A 6. Gently press the 5-Pos Variable Pathlength Cell Holder to connect the communication port (male) under the bottom of the 5-Pos Variable Pathlength Cell Holder to the port (female) in the sample compartment. Figure 9 Connecting the communication ports 7.
  • Page 7 09931462A M easurem ent Before executing System Self Test, the 5-Pos Variable Pathlength Cell Holder has to be installed first. NOTE: Otherwise, the instrument can be damaged electrically and do not operate properly. 1. Install the 5-Pos Variable Pathlength Cell Holder referring to chapter Installation. 2.
  • Page 8 09931462A 8. ① Click the Accessory tab and ② Select Variable Pathlength Multi-Cell on Accessory Type, and then click the ③Method Setup button to open the Multicell window. 9. The Multicell window will be shown. Using the button, select cells to be used among the available cells list.
  • Page 9 09931462A 13. Place the reference into the reference cell position and sample into the selected using cell positions. 14. Select the Sample button to measure. 15. After the measurement is finished, the results are displayed in the result window. Save or print results as required.
  • Page 10 09931462A 3. The functions of the Multicell calibration are shown as follows. Command Function MultiCell Position Show saved steps about each cell position of the Multi-Cell. Used for moving Multi-Cell position as clicking buttons Multicell Move Used for moving Multi-Cell position using buttons by entered step.
  • Page 11: Troubleshooting

     Recalibrate if the light beam does not go through the center of the cell holder hole. 2. Replace the lamp.  If the intensity value is still low after the recalibration, the lamps of the light source for measurement may be deteriorated and need replacing. Contact PerkinElmer.
  • Page 12 09931462A W hen the M ulti-Cell does not reset If the resetting is failed, the photo interrupter switch inside the 5-Pos Variable Pathlength Cell Holder needs replacing. Contact PerkinElmer for replacing the photo interrupter.