Installation Qualification Window

Use the Installation Qualification Window to perform the IQ test.
The IQ test verifies proper installation of the LabChip GX software
and verifies no unauthorized changes have been made to the
software. To open the Installation Qualification Window, select
Validation  Software IQ on the
The Installation Qualification Window contains the following
information:
Application
Verifies that the LabChip GX registry entries exist. If all expected
Information
registry entries are found, displays Passed. If any problems are
found, displays Failed.
Directory
Verifies that all folders exist. If all expected folders exist, displays
Structure
Passed. If any folders are not found, displays Failed.
Files
Verifies that all files that were installed by the LabChip GX
software still exist. If all file exist, displays Passed. If any files are
not found, displays Failed.
Summary
Displays Passed if all tests passed. Displays Failed if any of the
IQ tests failed.
Progress
Displays a progress bar while the IQ test is running.
View Details
Displays the detailed results of the current IQ. (Only displays after
Button
the IQ has been completed.)
Previous Result
Displays the results of the last IQ that was completed. (Only
button
available before another IQ test is run.)
Save As button
Saves the results of the IQ in an .xml file.
Start IQ button
Begins running the IQ.
Close button
Closes the Installation Qualification Window.
V4.2
Figure 92. Installation Qualification Window
LabChip GX User Manual