Changing the Time Window for Analysis

The Start Time and End Time parameters in the Peak Find tab
define the time window within which peaks are found.
To change the Start Time and End Time parameters for all wells in
the open assay:
1
2
3

Aligning or Unaligning the Marker Peaks

To perform data analysis for DNA, Protein, RNA, and Glycan
assays, the LabChip GX software aligns marker peaks included in
the sample wells with markers from the ladder.
The default setting for this function is enabled. Turning off the
analysis displays the data without aligning the markers in the wells
and the ladders.
V4.2
Select Analysis  Analysis Settings to open the
Analysis
Window, and then click the Peak Find Tab.
Change the parameters as necessary.
Click the OK button to save the setting, reanalyze the data, and
close the window.
Click the Apply button to apply the changes and reanalyze the
data, but keep the Assay Analysis window open.
To view the unaligned data (spike rejected, filtered and baseline
corrected), choose Analysis  Turn Off Analysis.
To re-enable analysis, choose Analysis  Turn On Analysis.
Figure 31. Data Before and After Alignment
LabChip GX User Manual
Data Analysis