[:INPut]:CURRent:EXTSensor:POJump?
Function
Queries the jump destination ranges of all
elements that are used when a current peak over-
range occurs.
Syntax
[:INPut]:CURRent:EXTSensor:POJump?
[:INPut]:CURRent:EXTSensor:POJump[:A
LL]
Function
Collectively sets the jump destination range of all
elements that is used when a current peak over-
range occurs.
Syntax
[:INPut]:CURRent:EXTSensor:
POJump[:ALL] {OFF|<Voltage>}
OFF = No jump destination current range
• When the crest factor is set to 3:
<Voltage> = 50 mV, 100 mV, 200 mV, 500 mV,
1 V, 2 V, 5 V, 10 V
• When the crest factor is set to 6 or 6A:
<Voltage> = 25 mV, 50 mV, 100 mV, 250 mV,
500 mV, 1 V, 2.5 V, 5 V
:INPUT:CURRENT:EXTSENSOR:POJUMP:
Example
ALL OFF
[:INPut]:CURRent:EXTSensor:POJump:EL
EMent<x>
Function
Sets or queries the jump destination range of the
specified element that is used when a current
peak over-range occurs.
[:INPut]:CURRent:EXTSensor:POJump:
Syntax
ELEMent<x> {OFF|<Voltage>}
[:INPut]:CURRent:EXTSensor:POJump:
ELEMent<x>?
<x> = 1 to 7 (element)
OFF = No jump destination current range
<Voltage> = See [:INPut]:CURRent:EXTSensor:
POJump[:ALL]
:INPUT:CURRENT:EXTSENSOR:POJUMP:
Example
ELEMENT1 10V
:INPUT:CURRENT:EXTSENSOR:POJUMP:
ELEMENT1?
-> :INPUT:CURRENT:EXTSENSOR:POJUMP:
ELEMENT1 10.00E+00
[:INPut]:CURRent:POJump?
Function
Queries the jump destination ranges of all
elements that are used when a current peak over-
range occurs.
[:INPut]:CURRent:POJump?
Syntax
IM WT5000-17EN
[:INPut]:CURRent:POJump[:ALL]
Function
Collectively sets the jump destination range of all
elements that is used when a current peak over-
range occurs.
Syntax
[:INPut]:CURRent:POJump[:ALL] {OFF|
<Current>}
OFF = No jump destination current range
760901 (30A High Accuracy Element)
• When the crest factor is set to 3:
<Current> = 500 mA, 1 A, 2 A, 5 A, 10 A, 20 A,
30 A
• When the crest factor is set to 6 or 6A:
<Current> = 250 mA, 500 mA, 1 A, 2.5 A, 5 A,
10 A, 15 A
760902 (5A High Accuracy Element)
• When the crest factor is set to 3:
<Current> = 5 mA, 10 mA, 20 mA, 50 mA,
100 mA, 200 mA, 500 mA, 1 A,
2 A, 5 A
• When the crest factor is set to 6 or
<Current> = 2.5 mA, 5 mA, 10 mA, 25 mA,
50 mA, 100 mA, 250 mA, 500 mA,
1 A, 2.5 A
:INPUT:CURRENT:POJUMP:ALL OFF
Example
[:INPut]:CURRent:POJump:ELEMent<x>
Function
Sets or queries the jump destination range of the
specified element that is used when a current
peak over-range occurs.
[:INPut]:CURRent:POJump:
Syntax
ELEMent<x> {OFF|<Current>}
[:INPut]:CURRent:POJump:ELEMent<x>?
<x> = 1 to 7 (element)
OFF = No jump destination current range
<Current> = See [:INPut]:CURRent:POJump[:ALL]
:INPUT:CURRENT:POJUMP:ELEMENT1 50A
Example
:INPUT:CURRENT:POJUMP:ELEMENT1?
-> :INPUT:CURRENT:POJUMP:
ELEMENT1 50.0E+00
[:INPut]:CURRent:RANGe?
Function
Queries the electric current ranges of all
elements.
[:INPut]:CURRent:RANGe?
Syntax

5.11 INPut Group

1
2
3
4
5
6A:
6
7
App
Index
5-49