Configuring the Instrument GP-IB Settings
This section explains the settings listed below. You must configure these settings when controlling the
instrument remotely through a GP-IB interface.
• GP-IB address
Remote Control Settings (Remote Control)
1.
Use either method below to open the utility settings overview screen.
• Procedure using the Setup menu: See section 1.4.
• Procedure using the UTILITY key: See section 1.4.
2.
Tap the Remote Control button.
The remote control setting screen (Network (VXI-11), GP-IB, USB (USB-TMC)) appears.
3.
Tap Address. Use the displayed input box to set the GP-IB address.
Remote Control button
Note
•
Only use one communication interface: GP-IB, USB, or Network. If you send commands simultaneously
from more than one communication interface, the instrument will not execute the commands properly.
•
When the controller is communicating with this instrument or with other devices through GP-IB, do not
change the address.
•
Each device that is connected by GP-IB has its own unique address in the GP-IB system. This address
is used to distinguish one device from other devices. Therefore, you must assign a unique address to the
instrument when connecting it to a PC or other device.