YOKOGAWA WT5000 User Manual page 66

Precision power analyzer, communication interface
Hide thumbs Also See for WT5000:
Table of Contents

Advertisement

:DISPlay:TRENd<x>:TDIV
Function
Sets or queries the trend horizontal axis (T/div).
:DISPlay:TRENd<x>:TDIV {<NRf>,<NRf>,
Syntax
<NRf>}
:DISPlay:TRENd<x>:TDIV?
<x> = 1 to 4 (trend display group)
{<NRf>,<NRf>,<NRf>} = 0,0,3 to 24,0,0
First <NRf> = 1, 3, 6, 12, 24 (hours)
Second <NRf> = 1, 3, 6, 10, 30 (minutes)
Third <NRf> = 3, 6, 10, 30 (seconds)
:DISPLAY:TREND1:TDIV 0,0,3
Example
:DISPLAY:TREND1:TDIV?
-> :DISPLAY:TREND1:TDIV 0,0,3
Description • This setting applies to the entire trend display
group. The <x> of TRENd<x> has no meaning
in the setting or query.
• Set the three <NRf>'s so that one is a non-zero
value and the other two are zero.
:DISPlay:TRENd<x>:VALue<x>
Function
Sets or queries the on/off status of the current
trend value display.
:DISPlay:TRENd<x>:
Syntax
VALue<x> {<Boolean>}
:DISPlay:TRENd<x>:VALue<x>?
<x> of VALue<x> = 1 or 2
1: Full screen or top screen
2: Bottom screen
:DISPLAY:TREND1:VALUE1 ON
Example
:DISPLAY:TREND1:VALUE1?
-> :DISPLAY:TREND1:VALUE1 1
Description This setting applies to the entire trend display
group. The <x> of TRENd<x> has no meaning in
the setting or query.
:DISPlay:VECTor<x>?
Function
Queries all vector display settings.
Syntax
:DISPlay:VECTor<x>?
<x> = 1 to 4 (vector display group)
:DISPlay:VECTor<x>:FORMat
Function
Sets or queries the display format of all vectors.
:DISPlay:VECTor<x>:FORMat {SINGle|
Syntax
DUAL}
:DISPlay:VECTor<x>:FORMat?
<x> = 1 to 4 (vector display group)
:DISPLAY:VECTOR1:FORMAT SINGLE
Example
:DISPLAY:VECTOR1:FORMAT?
-> :DISPLAY:VECTOR1:FORMAT SINGLE
:DISPlay:VECTor<x>:ITEM<x>?
Function
Queries all settings for the specified vector.
Syntax
:DISPlay:VECTor<x>:ITEM<x>?
<x> of VECTor<x> = 1 to 4 (vector display group)
<x> of ITEM<x> = 1 or 2 (item number)
IM WT5000-17EN
:DISPlay:VECTor<x>:ITEM<x>:OBJect
Function
Sets or queries the wiring unit that is displayed
using the specified vector.
:DISPlay:VECTor<x>:ITEM<x>:
Syntax
OBJect {<Element>}
:DISPlay:VECTor<x>:ITEM<x>:OBJect?
<x> of VECTor<x> = 1 to 4 (vector display group)
<x> of ITEM<x> = 1 or 2 (item number)
<Element> = {<NRf>|SIGMa|SIGMB|SIGMC}
(<NRf> = 1 to 7)
:DISPLAY:VECTOR1:ITEM1:OBJECT SIGMA
Example
:DISPLAY:VECTOR1:ITEM1:OBJECT?
-> :DISPLAY:VECTOR1:ITEM1:
OBJECT SIGMA
:DISPlay:VECTor<x>:ITEM<x>:{UMAG|IM
AG}
Function
Sets or queries the voltage or current zoom factor
for the vector display.
:DISPlay:VECTor<x>:ITEM<x>:{UMAG|
Syntax
IMAG} {<NRf>}
:DISPlay:VECTor<x>:ITEM<x>:{UMAG|
IMAG}?
<x> of VECTor<x> = 1 to 4 (vector display group)
<x> of ITEM<x> = 1 or 2 (item number)
<NRf> = 0.100 to 100.000
:DISPLAY:VECTOR1:ITEM1:UMAG 1
Example
:DISPLAY:VECTOR1:ITEM1:UMAG?
-> :DISPLAY:VECTOR1:ITEM1:UMAG 1.000
:DISPlay:VECTor<x>:NUMeric
Function
Sets or queries the on/off status of the numeric
data display on the vector display.
:DISPlay:VECTor<x>:
Syntax
NUMeric {<Boolean>}
:DISPlay:VECTor<x>:NUMeric?
<x> = 1 to 4 (vector display group)
Example
:DISPLAY:VECTOR1:NUMERIC ON
:DISPLAY:VECTOR1:NUMERIC?
-> :DISPLAY:VECTOR1:NUMERIC 1
5.6 DISPlay Group
1
2
3
4
5
6
7
App
Index
5-33

Advertisement

Table of Contents
loading

Table of Contents