YOKOGAWA WT5000 User Manual page 52

Precision power analyzer, communication interface
Hide thumbs Also See for WT5000:
Table of Contents

Advertisement

:COMMunicate:REMote
Function
Sets this instrument to remote or local mode. On
is remote mode.
:COMMunicate:REMote {<Boolean>}
Syntax
:COMMunicate:REMote?
Example
:COMMUNICATE:REMOTE ON
:COMMUNICATE:REMOTE?
-> :COMMUNICATE:REMOTE 1
:COMMunicate:VERBose
Function
Sets or queries whether the response to a
query is returned fully spelled out (example:
":INPUT:VOLTAGE:RANGE:ELEMENT1
1.000E+03") or in its abbreviated form (example:
"VOLT:RANG:ELEM1 1.000E+03").
:COMMunicate:VERBose {<Boolean>}
Syntax
:COMMunicate:VERBose?
Example
:COMMUNICATE:VERBOSE ON
:COMMUNICATE:VERBOSE?
-> :COMMUNICATE:VERBOSE 1
:COMMunicate:WAIT
Function
Waits for a specified extended event to occur.
:COMMunicate:WAIT <Register>
Syntax
<Register> = 0 to 65535 (Extended event register.
For more information, see page 6-5.)
:COMMUNICATE:WAIT 1
Example
Description For information about how to synchronize a
program using COMMunicate:WAIT, see page
4-9.
:COMMunicate:WAIT?
Function
Creates the response that is returned when a
specified extended event occurs.
:COMMunicate:WAIT? <Register>
Syntax
<Register> = 0 to 65535 (Extended event register.
For more information, see page 6-5.)
Example
:COMMUNICATE:WAIT? 65535 -> 1
IM WT5000-17EN
5.4 COMMunicate Group
5-19
1
2
3
4
5
6
7
App
Index