5.11 INPut Group
[:INPut]:FILTer[:LINE]:DLFilter:{NOR
mal|HARMonics}[:ALL]
Function
Collectively sets the digital line filter
{Normal|Harmonics} of all elements. (Advanced
settings)
Syntax
[:INPut]:FILTer[:LINE]:DLFilter:
{NORmal|HARMonics}[:ALL] {OFF|
<Frequency>}
OFF = Digital line filter off
<Frequency> = 0.1 kHz to 100.0 kHz (when the
digital line filter is on; cutoff frequency)
:INPUT:FILTER:LINE:DLFILTER:NORMAL:
Example
ALL OFF
Description This setting is valid when the advanced settings
([:INPut]:FILTer[:LINE]:ADVanced) are set to ON (1).
[:INPut]:FILTer[:LINE]:DLFilter:{NOR
mal|HARMonics}:ELEMent<x>
Function
Sets or queries the digital line filter
{Normal|Harmonics} of the specified element.
(Advanced settings)
[:INPut]:FILTer[:LINE]:DLFilter:
Syntax
{NORmal|HARMonics}:ELEMent<x> {OFF|
<Frequency>}
[:INPut]:FILTer[:LINE]:DLFilter:
{NORmal|HARMonics}:ELEMent<x>?
<x> = 1 to 7 (element)
OFF = Digital line filter off
<Frequency> = 0.1 kHz to 100.0 kHz (when the
digital line filter is on; cutoff frequency)
Example
:INPUT:FILTER:LINE:DLFILTER:NORMAL:
ELEMENT1 0.5KHZ
:INPUT:FILTER:LINE:DLFILTER:NORMAL:
ELEMENT1?
-> :INPUT:FILTER:LINE:DLFILTER:
NORMAL:ELEMENT1 500.0E+00
Description This setting is valid when the advanced settings
([:INPut]:FILTer[:LINE]:ADVanced) are set to ON (1).
[:INPut]:FILTer[:LINE]:TYPE
Function
Sets or queries the line filter type.
[:INPut]:FILTer[:LINE]:
Syntax
TYPE {BUTTerworth|BESSel}
[:INPut]:FILTer[:LINE]:TYPE?
:INPUT:FILTER:LINE:TYPE BUTTERWORTH
Example
:INPUT:FILTER:LINE:TYPE?
-> :INPUT:FILTER:LINE:
TYPE BUTTERWORTH
5-54
[:INPut]:FREQ2?
Function
Queries all second frequency measurement
settings.
[:INPut]:FREQ2?
Syntax
[:INPut]:FREQ2:HPFilter?
Function
Queries all high-pass filter settings (for second
frequency measurements).
[:INPut]:FREQ2:HPFilter?
Example
[:INPut]:FREQ2:HPFilter[:ALL]
Function
Collectively sets the high-pass filters (for second
frequency measurements) of all elements.
[:INPut]:FREQ2:HPFilter[:ALL] {OFF|
Syntax
<Frequency>}
OFF = High-pass filter off
<Frequency> = 0.1 Hz, 1 Hz, 10 Hz, 0.1 kHz to
100.0 kHz (when the high-pass filter is on;
cutoff frequency)
:INPUT:FREQ2:HPFILTER:ALL OFF
Example
[:INPut]:FREQ2:HPFilter:ELEMent<x>
Function
Sets or queries the high-pass filter (for second
frequency measurements) of the specified
element.
Syntax
[:INPut]:FREQ2:HPFilter:
ELEMent<x> {OFF|<Frequency>}
[:INPut]:FREQ2:HPFilter:ELEMent<x>?
<x> = 1 to 7 (element)
OFF = High-pass filter off
<Frequency> = 0.1 Hz, 1 Hz, 10 Hz, 0.1 kHz to
100.0 kHz (when the high-pass filter is on;
cutoff frequency)
:INPUT:FREQ2:HPFILTER:ELEMENT1 0.1HZ
Example
:INPUT:FREQ2:HPFILTER:ELEMENT1?
-> :INPUT:FREQ2:HPFILTER:
ELEMENT1 100.0E-03