Appendix 1 Error Messages
Code
Message
161
Invalid block data.
168
Block data not allowed.
171
Missing Right
172
Invalid expression.
178
Expression data not allowed.
181
Invalid outside macro definition.
Communication Execution Errors (200 to 299)
Code
Message
221
Setting conflict.
222
Data out of range.
223
Too much data.
224
Illegal parameter value.
225
Overflow.
226
Out Of Memory.
241
Hardware missing.
260
Expression error.
270
Macro error.
272
Macro execution error.
273
Illegal macro label.
275
Macro definition too long.
276
Macro recursion error.
277
Macro redefinition not allowed.
278
Macro header not found.
Communication Query Errors (400 to 499)
Code
Message
410
Query INTERRUPTED.
420
Query UNTERMINATED.
430
Query DEADLOCKED.
440
Query UNTERMINATED after indefinite
response.
System Communication Errors (300 and 399)
Code
Message
300
Communication device-specific error.
399
Fatal error in the communication driver.
App-2
Corrective Action
<Block data> cannot be used.
<Block data> cannot be used.
Equations cannot be used.
Equations cannot be used.
Equations cannot be used.
This instrument does not support the IEEE 488.2 macro
specifications.
Corrective Action
Check settings that are related to each other.
Check the ranges of the settings.
Check data byte lengths.
Check the ranges of the settings.
Keep program messages (including <PMT>) to 1024 bytes
or less in length.
Keep program messages (including <PMT>) to 1024 bytes
or less in length.
Check that the specified options are all installed.
Equations cannot be used.
This instrument does not support the IEEE 488.2 macro
specifications.
This instrument does not support the IEEE 488.2 macro
specifications.
This instrument does not support the IEEE 488.2 macro
specifications.
This instrument does not support the IEEE 488.2 macro
specifications.
This instrument does not support the IEEE 488.2 macro
specifications.
This instrument does not support the IEEE 488.2 macro
specifications.
This instrument does not support the IEEE 488.2 macro
specifications.
Corrective Action
Check the transmission and reception order.
Check the transmission and reception order.
Keep program messages (including <PMT>) to 1024 bytes
or less in length.
Do not write a query after *IDN? or *OPT?.
Corrective Action
Servicing required.
Servicing required.
Page
4-7 and
chapter 5
4-7 and
chapter 5
Chapter 5
Chapter 5
Page
Chapter 5
Chapter 5
Chapter 5
Chapter 5
4-2
4-2
Page
4-2
4-2
4-2
Page
IM WT5000-17EN