5.15 NUMeric Group
:NUMeric[:NORMal]:NUMber
Function
Sets or queries the number of numeric
data items that are transmitted by the
:NUMeric[:NORMal]:VALue? command.
:NUMeric[:NORMal]:NUMber {<NRf>|ALL}
Syntax
:NUMeric[:NORMal]:NUMber?
<NRf> = 1 to 1000 (ALL)
:NUMERIC:NORMAL:NUMBER 15
Example
:NUMERIC:NORMAL:NUMBER
-> :NUMERIC:NORMAL:NUMBER 15
Description • If the parameter is omitted from the :NUMeric
[:NORMal]:VALue? command, the numeric
data items from 1 to the specified value are
output in order.
• By default, the number of numeric data items is
set to 15.
:NUMeric[:NORMal]:PRESet
Function
Presets the numeric data output item pattern.
:NUMeric[:NORMal]:PRESet {<NRf>}
Syntax
<NRf> = 1 to 4
:NUMERIC:NORMAL:PRESET 1
Example
Description • For information about the output items that are
preset, see "Preset Patterns for Numeric Data
Output Items" at the end of this group on page
5-82.
• By default, the output items of Pattern 2 are
selected.
:NUMeric[:NORMal]:VALue?
Function
Queries the numeric data.
:NUMeric[:NORMal]:VALue? {<NRf>}
Syntax
<NRf> = 1 to 1000 (item number)
Example
• When <NRf> is specified:
:NUMERIC:NORMAL:VALUE? 1
-> 103.792E+00
• When <NRf> is omitted:
:NUMERIC:NORMAL:VALUE?
-> 103.792E+00,1.01438E+00,
105.285E+00,.. (omitted) ..,1.428E+00
• When :NUMeric:FORMat is set to FLOat:
:NUMERIC:NORMAL:VALUE?
-> #N (N-digit byte number)(data byte
sequence)
Description • If <NRf> is specified, only the numeric data for
the specified item is output.
• If <NRf> is omitted, the numeric data items
from 1 to the number specified by the :NUMeric
[:NORMal]:NUMber command are output in
order.
• For the formats of the individual numeric data
items that are output, see "Numeric Data
Formats" at the end of this group of commands
(page 5-81).