5.20 STORe Group
:STORe:NUMeric[:NORMal]:<Function>
Function
Sets or queries the on/off status of the specified
function's output when numeric data is stored.
:STORe:NUMeric[:NORMal]:
Syntax
<Function> {<Boolean>}
:STORe:NUMeric[:NORMal]:<Function>?
<Function> = {URMS|IRMS|P|S|Q|...}
:STORE:NUMERIC:NORMAL:URMS ON
Example
:STORE:NUMERIC:NORMAL:URMS?
-> :STORE:NUMERIC:NORMAL:URMS 1
Description For details on the <Function> options, see
"Numeric data functions" in "Function Option List
(Settings That Can Be Used for <Function>)" at
the end of the DISPlay group on page 5-36.
:STORe:NUMeric[:NORMal]:MOTor<x>
Function
Sets or queries the on/off status of the specified
motor's output when numeric data is stored.
:STORe:NUMeric[:NORMal]:
Syntax
MOTor<x> {<Boolean>}
:STORe:NUMeric[:NORMal]:MOTor<x>?
<x> = 1 to 4
:STORE:NUMERIC:NORMAL:MOTOR1 ON
Example
:STORE:NUMERIC:NORMAL:MOTOR1?
-> :STORE:NUMERIC:NORMAL:MOTOR1 1
:STORe:NUMeric[:NORMal]:PRESet<x>
Function
Presets the output on/off pattern of the element
functions to be used when numeric data is stored.
:STORe:NUMeric[:NORMal]:PRESet<x>
Syntax
<x> = 1 or 2 (preset number)
:STORE:NUMERIC:NORMAL:PRESET1
Example
Description For details on the storage item setting patterns
that result when the pattern is reset, see the
features guide, IM WT5000-01EN.
:STORe:PAUSe
Function
Pauses the storing of numeric data.
:STORe:PAUSe
Syntax
:STORE:PAUSE
Example
Description The conventional command ":STORe:STOP" can
also be used.
:STORe:RECord
Function
Starts the storing of numeric data.
Syntax
:STORe:RECord
:STORE:RECORD
Example
Description • If :STORe:SMODe is set to MANual, the storage
operation is executed. If :STORe:SMODe
is set to RTIMe, INTEGrate, or EVENt, this
instrument enters into a storage wait state. If
:STORe:SMODe is set to SINGle, the storage
operation is executed.
• The conventional command ":STORe:STARt"
can also be used.
5-88
:STORe:RTIMe?
Function
Queries the storage start and end times for real-
time storage mode.
:STORe:RTIMe?
Syntax
:STORe:RTIMe:{STARt|END}
Function
Sets or queries the storage start or end time for
real-time storage mode.
:STORe:RTIMe:{STARt|END} {<NRf>,
Syntax
<NRf>,<NRf>,<NRf>,<NRf>,<NRf>}
:STORe:RTIMe:{STARt|END}?
{<NRf>,<NRf>,<NRf>,<NRf>,<NRf>,<NRf>} =
2001,1,1,0,0,0 to 2099,12,31,23,59,59
First <NRf> = 2001 to 2099 (year)
Second <NRf> = 1 to 12 (month)
Third <NRf> = 1 to 31 (day)
Fourth <NRf> = 0 to 23 (hour)
Fifth <NRf> = 0 to 59 (minute)
Sixth <NRf> = 0 to 59 (second)
:STORE:RTIME:START 2018,1,1,0,0,0
Example
:STORE:RTIME:START?
-> :STORE:RTIME:START 2018,1,1,0,0,0
Description This command is valid when the storage mode
(:STORe:SMODe) is set to RTIMe.
:STORe:SASTart
Function
Sets or queries whether numeric data is stored
when storage starts.
:STORe:SASTart {<Boolean>}
Syntax
:STORe:SASTart?
Example
:STORE:SASTART OFF
:STORE:SASTART? -> :STORE:SASTART 0
Description This command is valid when the storage mode
(:STORe:SMODe) is set to MANual, RTIMe, or
INTEGrate (and when the storage interval is a
value other than 0 for MANual or RTIMe mode).
:STORe:SMODe
Function
Sets or queries the storage mode.
:STORe:SMODe {MANual|RTIMe|
Syntax
INTEGrate|EVENt|SINGle}
:STORe:SMODe?
MANual = Manual storage mode
RTIMe = Real-time storage mode
INTEGrate = Integration-synchronized storage
mode
EVENt = Event-synchronized storage mode
SINGle = Single-shot storage mode
:STORE:SMODE MANUAL
Example
:STORE:SMODE? -> :STORE:SMODE MANUAL