Cursor Group - YOKOGAWA WT5000 User Manual

Precision power analyzer, communication interface
Hide thumbs Also See for WT5000:
Table of Contents

Advertisement

5.5

CURSor Group

The commands in this group deal with cursor measurements.
You can make the same settings and queries that you can on the screen that appears by using Setup menu >
Computation/Output tab > Display button > Cursors button.
:CURSor?
Function
Queries all cursor measurement settings.
:CURSor?
Syntax
:CURSor:BAR<x>?
Function
Queries all bar graph display cursor measurement
settings.
:CURSor:BAR<x>?
Syntax
<x> = 1 to 4 (bar graph display group)
:CURSor:BAR<x>:LINKage
Function
Sets or queries the on/off status of the cursor
position linkage on the bar graph display.
:CURSor:BAR<x>:LINKage {<Boolean>}
Syntax
:CURSor:BAR<x>:LINKage?
<x> = 1 to 4 (bar graph display group)
:CURSOR:BAR:LINKAGE OFF
Example
:CURSOR:BAR:LINKAGE?
-> :CURSOR:BAR1:LINKAGE 0
:CURSor:BAR<x>:POSition<x>
Function
Sets or queries the position of the specified
cursor on the bar graph display.
:CURSor:BAR<x>:POSition<x> {<NRf>}
Syntax
:CURSor:BAR<x>:POSition<x>?
<x> of BAR<x> = 1 to 4 (bar graph display group)
<x> of POSition<x> = 1, 2 (1 = C1 +, 2 = C2 x)
<NRf> = 0 to 500
:CURSOR:BAR:POSITION1 1
Example
:CURSOR:BAR:POSITION1?
-> :CURSOR:BAR1:POSITION1 1
:CURSor:BAR<x>[:STATe]
Function
Sets or queries the on/off status of the cursor
display on the bar graph display.
Syntax
:CURSor:BAR<x>[:STATe] {<Boolean>}
:CURSor:BAR<x>:STATe?
<x> = 1 to 4 (bar graph display group)
:CURSOR:BAR:STATE ON
Example
:CURSOR:BAR:STATE?
-> :CURSOR:BAR1:STATE 1
5-20
:CURSor:BAR<x>:{Y<x>|DY}?
Function
Queries the measured value of the specified
cursor on the bar graph display.
:CURSor:BAR<x>:{Y<x>|DY}?
Syntax
<x> of BAR<x> = 1 to 4 (bar graph display group)
Y<x> = Y-axis value of the cursor position
(Y1 = Y1 +, Y2 +, Y3 +; Y2 = Y1 x, Y2 x, Y3 x)
DY = Difference between the cursor Y-axis values
(DY1, DY2, DY3)
:CURSOR:BAR:Y1? -> 78.628E+00
Example
Description • When multiple bar graphs are displayed, the
cursor measured values of each bar graph are
returned in order.
• If the bar graph cursor display is not turned on,
NAN (Not A Number) is returned.
:CURSor:TRENd<x>?
Function
Queries all trend display cursor measurement
settings.
:CURSor:TRENd<x>?
Syntax
<x> = 1 to 4 (trend display group)
:CURSor:TRENd<x>:LINKage
Function
Sets or queries the on/off status of the cursor
position linkage on the trend display.
Syntax
:CURSor:TRENd<x>:LINKage {<Boolean>}
:CURSor:TRENd<x>:LINKage?
<x> = 1 to 4 (trend display group)
:CURSOR:TREND:LINKAGE OFF
Example
:CURSOR:TREND:LINKAGE?
-> :CURSOR:TREND1:LINKAGE 0
:CURSor:TRENd<x>:POSition<x>
Function
Sets or queries the position of the specified
cursor on the trend display.
Syntax
:CURSor:TRENd<x>:POSition<x> {<NRf>}
:CURSor:TRENd<x>:POSition<x>?
<x> of TRENd<x> = 1 to 4 (trend display group)
<x> of POSition<x> = 1, 2 (1 = C1 +, 2 = C2 x)
<NRf> = 0 to 2001
Example
:CURSOR:TREND:POSITION1 200
:CURSOR:TREND:POSITION1?
-> :CURSOR:TREND1:POSITION1 200
IM WT5000-17EN

Advertisement

Table of Contents
loading

Table of Contents