Input Group - YOKOGAWA WT310EH User Manual

Digital power meter communication interface
Hide thumbs Also See for WT310EH:
Table of Contents

Advertisement

6.7

INPut Group

The commands in this group deal with the measurement conditions of the input elements.
You can make the same settings and queries that you can by pressing the WIRING, MODE, VOLTAGE, CURRENT,
or UTILITY key on the front panel and then using the CF or rAnGE menu or by pressing the SETUP key and then
using the SCALE, rAtio, SYnC, L.FiLt, or F.FiLt menu.
:INPut?
Function
Queries all input settings.
Syntax
:INPut?
[:INPut]:CFACtor
Function
Sets or queries the crest factor.
Syntax
[:INPut]:CFACtor {<NRf>|A6}
[:INPut]:CFACtor?
<NRf> = 3 , 6
A6 = Display range expand mode (6A) for crest
factor 6
Example
:INPUT:CFACTOR 3
:INPUT:CFACTOR?
-> :INPUT:CFACTOR 3
[:INPut]:WIRing
Function
Sets or queries the wiring system.
Syntax
[:INPut]:WIRing {(P1W2|P1W3|P3W3|
P3W4|V3A3)}
[:INPut]:WIRing?
P1W2 = Single-phase, two-wire system [1P2W]
P1W3 = Single-phase, three-wire system
[1P3W]
P3W3 = Three-phase, three-wire system [3P3W]
P3W4 = Three-phase, four-wire system [3P4W]
V3A3 = Three-phase, three-wire system with a
three-voltage, three-current method [3V3A]
Example
:INPUT:WIRING P1W3
:INPUT:WIRING?
-> :INPUT:WIRING P1W3
Description • For the WT310E and WT310EH, the wiring
system is fixed to P1W2. No other setting is
allowed.
• For the WT332E and WT333E, the wiring
system cannot be set to P1W2.
[:INPut]:MODE
Function
Sets or queries the voltage and current
measurement mode.
Syntax
[:INPut]:MODE {RMS|VMEan|DC}
[:INPut]:MODE?
Example
:INPUT:MODE RMS
:INPUT:MODE? -> :INPUT:MODE RMS
[:INPut]:VOLTage?
Function
Queries all voltage measurement settings.
Syntax
[:INPut]:VOLTage?
6-12
[:INPut]:VOLTage:RANGe
Function
Sets or queries the voltage range.
Syntax
[:INPut]:VOLTage:RANGe {< Voltage >}
[:INPut]:VOLTage:RANGe?
• When the crest factor is set to 3
<Voltage> = 15, 30, 60, 150, 300, 600(V)
• When the crest factor is set to 6 or 6A
<Voltage> = 7.5, 15, 30, 75, 150, 300(V)
Example
:INPUT:VOLTAGE:RANGE 600V
:INPUT:VOLTAGE:RANGE?
-> :INPUT:VOLTAGE:RANGE 600.0E+00
[:INPut]:VOLTage:AUTO
Function
Sets or queries the voltage auto range on/off
state.
Syntax
[:INPut]:VOLTage:AUTO {<Boolean>}
[:INPut]:VOLTage:AUTO?
Example
:INPUT:VOLTAGE:AUTO ON
:INPUT:VOLTAGE:AUTO?
-> :INPUT:VOLTAGE:AUTO 1
[:INPut]:VOLTage:CONFig
Function
Sets or queries the valid voltage range.
Syntax
[:INPut]:VOLTage:CONFig
{ALL|< Voltage >[,< Voltage >]
[,< Voltage >]...}
[:INPut]:VOLTage:CONFig?
ALL = All ranges are valid.
<Voltage> = See ( :INPut:VOLTage:
RANGe ).
Example
:INPUT:VOLTAGE:CONFIG ALL
:INPUT:VOLTAGE:CONFIG?
-> :INPUT:VOLTAGE:CONFIG ALL
:INPUT:VOLTAGE:CONFIG 600,150,15
:INPUT:VOLTAGE:CONFIG?
-> :INPUT:VOLTAGE:CONFIG 600.0E+00,
150.0E+00,15.0E+00
Description In the parameters, list the voltage ranges that you
want to enable. To enable all the ranges, specify
the parameter "ALL."
IM WT310E-17EN