5.6 DISPlay Group
:DISPlay:BAR<x>:ITEM<x>:SCALing:VAL
ue
Function
Sets or queries the upper limit of the manual
scaling of the specified bar graph.
:DISPlay:BAR<x>:ITEM<x>:SCALing:
Syntax
VALue {<NRf>}
:DISPlay:BAR<x>:ITEM<x>:SCALing:
VALue?
<x> of BAR<x> = 1 to 4 (bar graph display group)
<x> of ITEM<x> = 1 to 3 (item number)
<NRf> = 0 to 9.999E+12
:DISPLAY:BAR1:ITEM1:SCALING:
Example
VALUE 100
:DISPLAY:BAR1:ITEM1:SCALING:VALUE?
-> :DISPLAY:BAR1:ITEM1:SCALING:
VALUE 100.0E+00
Description • This command is valid when the scaling mode
of the bar graph (:DISPlay:BAR<x>:ITEM<x>:
SCALing:MODE) is set to MANual.
• This command only sets the upper limit. The
lower limit is determined automatically (as
shown below) according to the vertical scaling
mode (:DISPlay:BAR<x>:ITEM<x>:SCALing:
VERTical).
- When the mode is LINear: 0 when the X-axis
position (:DISPlay:BAR<x>:ITEM<x>:SCAL
ing:XAXis) is "BOTTom;" the negative value
of the upper limit when the X-axis position is
"CENTer"
- When the mode is LOG: The upper
limit/10000
:DISPlay:BAR<x>:ITEM<x>:SCALing:VERT
ical
Function
Sets or queries the vertical scaling mode of the
specified bar graph.
:DISPlay:BAR<x>:ITEM<x>:SCALing:
Syntax
VERTical {LINear|LOG}
:DISPlay:BAR<x>:ITEM<x>:SCALing:
VERTical?
<x> of BAR<x> = 1 to 4 (bar graph display group)
<x> of ITEM<x> = 1 to 3 (item number)
:DISPLAY:BAR1:ITEM1:SCALING:
Example
VERTICAL LOG
:DISPLAY:BAR1:ITEM1:SCALING:
VERTICAL?
-> :DISPLAY:BAR1:ITEM1:SCALING:
VERTICAL LOG
Description This command is valid when the scaling mode of
the bar graph (:DISPlay:BAR<x>:ITEM<x>:SCALi
ng:MODE) is set to MANual.
5-24
:DISPlay:BAR<x>:ITEM<x>:SCALing:XAX
is
Function
Sets or queries the position of the X axis of the
specified bar graph.
:DISPlay:BAR<x>:ITEM<x>:SCALing:
Syntax
XAXis {BOTTom|CENTer}
:DISPlay:BAR<x>:ITEM<x>:SCALing:
XAXis?
<x> of BAR<x> = 1 to 4 (bar graph display group)
<x> of ITEM<x> = 1 to 3 (item number)
Example
:DISPLAY:BAR1:ITEM1:SCALING:
XAXIS BOTTOM
:DISPLAY:BAR1:ITEM1:SCALING:XAXIS?
-> :DISPLAY:BAR1:ITEM1:SCALING:
XAXIS BOTTOM
Description This command is valid when the scaling mode of
the bar graph (:DISPlay:BAR<x>:ITEM<X>:
SCALing:MODE) is set to MANual and the vertical
scaling mode of the bar graph (:DISPlay:BAR<x>:
ITEM<x>:SCALing:VERTical) is set to LINear.
:DISPlay:BAR<x>:ORDer
Function
Sets or queries the displayed starting and ending
harmonic orders of the bar graphs.
:DISPlay:BAR<x>:ORDer {<NRf>,<NRf>}
Syntax
:DISPlay:BAR<x>:ORDer?
<x> = 1 to 4 (bar graph display group)
First <NRf> = 0 to 490 (displayed starting
harmonic order)
Second <NRf> = 10 to 500 (displayed ending
:DISPLAY:BAR1:ORDER 1,100
Example
:DISPLAY:BAR1:ORDER?
-> :DISPLAY:BAR1:ORDER 1,100
Description • Set the starting harmonic order and then the
ending harmonic order.
• Set the ending harmonic order to a value
greater than or equal to that of the starting
harmonic order + 10.
harmonic order)