YOKOGAWA WT5000 User Manual page 102

Precision power analyzer, communication interface
Hide thumbs Also See for WT5000:
Table of Contents

Advertisement

:MEASure:MHOLd
Function
Sets or queries the on/off status of the MAX
HOLD feature used in user-defined functions.
:MEASure:MHOLd {<Boolean>}
Syntax
:MEASure:MHOLd?
Example
:MEASURE:MHOLD ON
:MEASURE:MHOLD? -> :MEASURE:MHOLD 1
Description • The MAX HOLD operation starts when the
MAX HOLD feature is specified by a user-
defined function and :MEASure:MHOLd is set
to ON.
• When :MEASure:MHOLd is set to OFF, the
MAX HOLD operation ends, and the MAX
HOLD value becomes "no data."
• If :MEASure:MHOLd is set to ON after having
already been set to ON before, the MAX HOLD
value is reset, and the MAX HOLD operation
starts again.
• For information about specifying the MAX
HOLD feature, see the features guide, IM
WT5000-01EN.
:MEASure:PC?
Function
Queries all Pc (Corrected Power) computation
settings.
Syntax
:MEASure:PC?
:MEASure:PC:FORMula
Function
Sets or queries the Pc (Corrected Power)
equation.
:MEASure:PC:FORMula {TYPE1|TYPE2}
Syntax
:MEASure:PC:FORMula?
Example
:MEASURE:PC:FORMULA TYPE1
:MEASURE:PC:FORMULA?
-> :MEASURE:PC:FORMULA TYPE1
Description For details on the equations available for TYPE1
or TYPE2, see the features guide, IM W5000-
01EN.
:MEASure:PC:P<x>
Function
Sets or queries a Pc (Corrected Power) equation
parameter.
:MEASure:PC:P<x> {<NRf>}
Syntax
:MEASure:PC:P<x>?
<x> = 1, 2 (P1, P2)
<NRf> = 0.0001 to 9.9999
:MEASURE:PC:P1 0.5
Example
:MEASURE:PC:P1?
-> :MEASURE:PC:P1 0.5000
Description These parameters are used when :MEASure:PC:
FORMula is set to TYPE1.
IM WT5000-17EN
:MEASure:PHASe
Function
Sets or queries the display format of the phase
difference.
:MEASure:PHASe {<NRf>}
Syntax
:MEASure:PHASe?
<NRf> = 180 or 360
:MEASURE:PHASE 180
Example
:MEASURE:PHASE?
-> :MEASURE:PHASE 180
Description When 180 is selected, the phase is displayed
using 0 to ±180° (Lead/Lag). When 360 is
selected, the phase is displayed using 0° to 360°.
:MEASure:SFORmula
Function
Sets or queries the equation used to compute S
(apparent power).
:MEASure:SFORmula {RMS|MEAN|DC|
Syntax
MRMS|RMEAN}
:MEASure:SFORmula?
:MEASURE:SFORMULA RMS
Example
:MEASURE:SFORMULA?
-> :MEASURE:SFORMULA RMS
Description The equations that correspond to each option are
as follows:
RMS:
S = Urms*Irms
MEAN:
S = Umean*Imean
DC:
S = Udc*Idc
MRMS: S = Umean*Irms
RMEAN: S = Urmean*Irmean
:MEASure:SQFormula
Function
Sets or queries the equation used to compute S
(apparent power) and Q (reactive power).
:MEASure:SQFormula {TYPE1|TYPE2|
Syntax
TYPE3}
:MEASure:SQFormula?
:MEASURE:SQFORMULA TYPE1
Example
:MEASURE:SQFORMULA?
-> :MEASURE:SQFORMULA TYPE1
Description For details on the equations available for TYPE1,
TYPE2, or TYPE3, see the features guide, IM
W5000-01EN.
:MEASure:SYNChronize
Function
Sets or queries the synchronized measurement
mode.
:MEASure:SYNChronize {MASTer|SLAVe}
Syntax
:MEASure:SYNChronize?
:MEASURE:SYNCHRONIZE MASTER
Example
:MEASURE:SYNCHRONIZE?
-> :MEASURE:SYNCHRONIZE MASTER
5.13 MEASure Group
Index
5-69
1
2
3
4
5
6
7
App