YOKOGAWA WT5000 User Manual page 90

Precision power analyzer, communication interface
Hide thumbs Also See for WT5000:
Table of Contents

Advertisement

[:INPut]:RLINkage
Function
Sets or queries the on/off status of the
measurement range link by wiring unit Σ.
[:INPut]:RLINkage {<Boolean>}
Syntax
[:INPut]:RLINkage?
Example
:INPUT:RLINKAGE ON
:INPUT:RLINKAGE?
-> :INPUT:RLINKAGE 1
Description This is valid only when there are two or more
installed input elements.
[:INPut]:SCALing?
Function
Queries all scaling settings.
[:INPut]:SCALing?
Syntax
[:INPut]:SCALing:CTPReset?
Function
Queries the CT ratio presets of all elements.
Syntax
[:INPut]:SCALing:CTPReset?
[:INPut]:SCALing:CTPReset[:ALL]
Function
Collectively sets the CT ratio presets of all
elements.
[:INPut]:SCALing:
Syntax
CTPReset[:ALL] {OTHERS|CT2000A|
CT1000|CT200|CT60}
Example
:INPUT:SCALING:CTPRESET:ALL CT1000
[:INPut]:SCALing:CTPReset:ELEMent<x>
Function
Sets or queries the CT ratio preset of the
specified element.
[:INPut]:SCALing:CTPReset:
Syntax
ELEMent<x> {OTHERS|CT2000A|CT1000|
CT200|CT60}
[:INPut]:SCALing:CTPReset:ELEMent<x>?
<x> = 1 to 7 (element)
:INPUT:SCALING:CTPRESET:
Example
ELEMENT1 CT1000
Description If a preset is queried when the preset value and
the CT ratio value are different, "AST" is added to
the preset name.
[:INPut]:SCALing:CTPReset:{SIGMA|SIG
MB|SIGMC}
Function
Collectively sets the CT ratio presets of all the
elements that belong to the specified wiring unit
(ΣA, ΣB, or ΣB).
[:INPut]:SCALing:CTPReset:{SIGMA|
Syntax
SIGMB|SIGMC} {OTHERS|CT2000A|
CT1000|CT200|CT60}
:INPUT:SCALING:CTPRESET:SIGMA CT1000
Example
Description SIGMA, SIGMB, or SIGMC is invalid if the wiring
system setting ([:INPut]:WIRing) is made in such
a way that the corresponding wiring unit (ΣA, ΣB,
or ΣC) does not exist.
IM WT5000-17EN
[:INPut]:SCALing:STATe?
Function
Queries the on/off statuses of the scaling of all
elements.
[:INPut]:SCALing:STATe?
Syntax
[:INPut]:SCALing[:STATe][:ALL]
Function
Collectively sets the on/off status of the scaling of
all elements.
[:INPut]:SCALing[:STATe][:
Syntax
ALL] {<Boolean>}
:INPUT:SCALING:STATE:ALL OFF
Example
[:INPut]:SCALing[:STATe]:ELEMent<x>
Function
Sets or queries the on/off status of the scaling of
the specified element.
[:INPut]:SCALing[:STATe]:
Syntax
ELEMent<x> {<Boolean>}
[:INPut]:SCALing[:STATe]:ELEMent<x>?
<x> = 1 to 7 (element)
:INPUT:SCALING:STATE:ELEMENT1 OFF
Example
:INPUT:SCALING:STATE:ELEMENT1?
-> :INPUT:SCALING:STATE:ELEMENT1 0
[:INPut]:SCALing:{VT|CT|SFACtor}?
Function
Queries the VT ratios, CT ratios, or power
coefficients of all elements.
[:INPut]:SCALing:{VT|CT|SFACtor}?
Syntax
[:INPut]:SCALing:{VT|CT|SFACtor}
[:ALL]
Function
Collectively sets the VT ratio, CT ratio, or power
coefficient of all elements.
Syntax
[:INPut]:SCALing:{VT|CT|
SFACtor}[:ALL] {<NRf>}
<NRf> = 0.0001 to 99999.9999
:INPUT:SCALING:VT:ALL 1
Example
[:INPut]:SCALing:{VT|CT|SFACtor}:ELE
Ment<x>
Function
Sets or queries the VT ratio, CT ratio, or power
coefficient of the specified element.
[:INPut]:SCALing:{VT|CT|SFACtor}:
Syntax
ELEMent<x> {<NRf>}
[:INPut]:SCALing:{VT|CT|SFACtor}:
ELEMent<x>?
<x> = 1 to 7 (element)
<NRf> = 0.0001 to 99999.9999
:INPUT:SCALING:VT:ELEMENT1 1
Example
:INPUT:SCALING:VT:ELEMENT1?
-> :INPUT:SCALING:VT:ELEMENT1 1.0000
5.11 INPut Group
1
2
3
4
5
6
7
App
Index
5-57