8.2

GP-IB Interface Functions and Specifications

GP-IB Interface Functions
Listener function
• All of the settings that you can enter with the
panel keys can be set through the GP-IB
interface except for turning ON/OFF the power
and setting the communication parameters.
• Receives commands from a controller requesting
the output of setup information, measured data,
and other information.
• Also receives status report commands.
Talker function
Outputs setup information, measured data, and
other information.
Note
Talk-only, listen-only, and controller functions are not available
on this instrument.
Switching between Remote and Local
Modes
When switching from local to remote mode
Receiving a REN (Remote Enable) message from
the controller when the instrument is in the local
mode causes the instrument to switch to the remote
mode.
• The REMOTE indicator turns ON (see page 2-1).
• All keys other than the LOCAL key are locked.
• The settings that existed in the local mode are
maintained even when the instrument switches to
the remote mode.
When switching from remote to local mode
Pressing the LOCAL key when the instrument is in
the remote mode causes the instrument to switch to
the local mode. However, this act is invalid if the
instrument has been set to Local Lockout mode
(see next page) by the controller.
• The REMOTE indicator turns OFF.
• Key operations are enabled.
• The settings that existed in the remote mode are
maintained even when the instrument switches to
the local mode.
IM 704420-01E
GP-IB Interface Specifications
• Electrical and mechanical specifications
Conforms to IEEE St'd 488-1978
• Functional specifications
See the table below
• Code
ISO (ASCII) code
• Mode
Addressable mode
• Address setting
The address can be set in the range from 0 to 30
using the GP-IB address setting that is displayed
with the SHIFT+LOCAL key.
• Clear remote mode
Remote mode can be cleared by pressing the
LOCAL key except when the instrument has been
set to Local Lockout mode by the controller.
Function
Subset Name
Source handshaking
SH1
Acceptor handshaking
AH1
Talker
T6
Listener
L4
Service request
SR1
Remote local
RL1
Parallel polling
PP0
Clear device
DC1
Device trigger
DT1
Controller
C0
Electrical characteristics E1
Description
Full source
handshaking capability
Full acceptor
handshaking capability
Basic talker capability,
8
serial polling, untalk on
MLA (My Listen
, and no talk-
Address)
only capability
Basic listener capability,
unlisten on
MTA (My
, and no
Talk Address)
listen-only capability.
Full service request
capability
Full remote/local
capability
No parallel polling
capability
Full device clear
capability
Full device trigger
capability
No controller functions
Open collector
8-3