Technical Specifications

Performance specifications

Performance
specifications
specifications
In conformity
requirements
64
Thermo Scientific KingFisher Duo User Manual
The performance specifications are presented in Table 6–6.
Table 6–6. Performance specifications
Performance specifications
Processing volume
Capacity (samples per run)
Collection efficiency of the
particles (indoor use)
Magnetic particle size
Magnet rods
Plates per deck
Plate types (disposable)
* Recommended filling volume
Elution strip
Tip combs (polypropylene –
disposable)
Automatic tip load
Heating/cooling temperature
- plate row block
- elution strip block
Heating block accuracy
Keypad / Display
Safety
This section describes the safety specifications for the KingFisher Duo
instrument.
KingFisher Duo bears the following markings:
with the
Type 706
+15 VDC / 6 A
CE mark
50–1000 μl (96 DW plate, 12-pin magnet head)
30–130 μl (elution strip, 12-pin magnet head)
200–5000 μl (24 DW plate, 6-pin magnet head)
Up to 12 (12-pin magnet head)
Up to 6 (6-pin magnet head)
≥ 95%, Microtiter deep well 96 plate, neutral wash buffer
containing detergent, 2.8 μm particles, 3 collections, RT
ca. > 1 μm
12 or 6, interchangeable KingFisher Duo heads
2
Microtiter deep well 96 plate (50–1000 μl*)
KingFisher Flex 24 deep well plate (200–5000 μL*)
Special design, 1 x 12 format
12 / 6 in one frame for Microtiter deep well 96 plate or KingFisher
Flex 24 deep well plate
Yes
From +10°C to 75°C, instrument in RT
From 4°C to 75°C, instrument in RT
± 1°C, up to +75°C, instrument in RT
/
/
/
START
PAUSE
STOP
OK
(Left arrow key) / LCD 5" 480 x 800 pixel color display
/
/ four arrow keys &