Chapter 13 CD measurements using phase shift method

Performing a CD measurement through amplifiers

Performing a CD measurement through
amplifiers
The CD reference step is not required for measuring through amplifiers:
1
Connect the source and the 81DISPAP/81MRDISPAP module at
each end of the fiber under test. Optical attenuators (OLA55 1 and
OLA55 2) have to be positioned before and after the amplified link.
Viavi recommends to:
setup the front attenuator (OLA55 #1) so that the power level
received by the amplifier is 10 dB lower than the in-service power
level (in normal operation).
setup the far end attenuator (OLA55 #2) so that the maximum
received power level by the CD tester is -10dBm max
81BBSXX or
OBS-5XX
2
Press the S
3
Select the appropriate CD test setup according to your application as
earlier defined in
4
Select Yes for Amplified link (It will limit the acquisition band from
1530 nm to 1565 nm).
5
Press the S
NOTE
Depending on the amplifier specification, it may be required to
adjust the acquisition band to a short wavelength range.
454
Amplified Network
OLA 55 #1
Psource=Pi - 10dB
button to access the CD configuration menu.
ETUP
"Configuring the CD test" on page
/S
TART
TOP
User Manual
OLA 55 #2
Ptester=- 10dBm max
button to perform CD measurement.
78000010220 Rev. 023
81DISPAP or
81MRDISPAP
440.