Advertisement

Chapter 8 Optical Spectrum Measurement
EDFA results analysis
Other S
ETUP
ments. Refer to
complete description.
NOTE
If your Platform 8000 is equipped with an OSA-303, you may use
the two ports to test before and after the EDFA. In this case,
make sure you select «Port A+B» for

EDFA measurements

Measurement procedure, using one port :
If only one port is selected, the Platform 8000 is ready to perform the «Acq.
in» (signal before being amplified by EDFA).
1
Connect your Platform 8000 to your fiber before the EDFA.
2
Click S
3
Switch to
4
Connect your Platform 8000 to your fiber after the EDFA.
5
Click S
Results appear automatically in the table.
316
parameters are the same for EDFA as for WDM measure-
"OSA optical spectrum analyzer Setup" on page 267
/S
to perform the first acquisition.
TART
TOP
Acq.
Out.
/S
to perform the second acquisition.
TART
TOP
User Manual
Input
78000010220 Rev. 023
Port.
for a