Figure 67
SmartLink function
Zone 1
Zone 2
Zone 3
The screen is divided into three zones:
•
Zone 1: Graphical representation of the link, with icons symbolizing
the different events detected.
•
Zone 2: Link Table, which gives a summary of results for each wave-
length, with results within/without thresholds in green/red (according
to Alarm thresholds defined in the setup screen).
•
Zone 3: Alarms table (if any).
CAUTION
If several traces are displayed in overlay, with the same
wavelength, then only the Zone 2 is displayed, there is no
graphical representation of the link (Zone 1).
Getting a diagnostic on potential causes of failure
When an event has reached an alarm, the diagnostic of potential causes is
given in SmartLink view.
1
Select the event to be described (underlined in yellow).
User Manual
Chapter 3 Reflectometry measurements
78000010220 Rev. 023
FTTA-SLM option