Chapter 16 Fiber Characterization
Fiber Characterization Automated Test Sequence using the Link Characterization Script
Figure 228 OTDR Alarms - Default and User
Selecting data storage directory
1
Press the F
2
Press File/Explorer softkey to select the "Explorer" page.
3
Using the navigation keys or touchscreen, on the left side of the
screen, select the main or sub directory where you wish to store files,
typically the "Hard Disk".
4
If necessary, create a new directory to store the results: see
"Creating the storage directory" on page
Launching the Link Characterization script
1
Press the S
2
Select the function Script > Link Characterization using touch-
screen or stylus.
Figure 229 Script - Link Characterization function selected
3
Press the soft key Launch to enter the script configuration menu.
512
hardkey.
ILE
hard key on the 8000 V2 and 6000A units
CRIPT
User Manual
494.
78000010220 Rev. 023