Chapter 11 High Resolution OSA Measurement
Trace display functions
2
Select the key
Each click on this key will alternatively insert or delete the check mark
against
Full scale
To display the entire trace, with no zoom or displacement:
1
Either press the
or,
With
Shift function
The Shift function is used to displace the displayed section of the trace by
pressing the direction keys.
The horizontal shift is performed maintaining the point of intersection
between the trace and the selected cursor at the same level, scrolling the
trace horizontally while following it vertically, so that it never goes off the
screen.
To use this function:
1
Select the zoom factor as described above.
2
Choose cursor and cursor position.
3
On the
4
Use the direction keys or touch and hold screen to shift the trace in
the desired direction.
Trace /Table key
This key offers a choice from the following displays:
Trace alone: main display of the trace with a single line of the table at the
foot of the page
398
CursorX/CursorY
Cursor Y
Full Scale
Trace
selected on Trace/Channel, press validation key
Zoom/Shift
key, select Shift.
(see"Example of OSA test result (with grid)" on page
User Manual
to modify the current choice.
key
78000010220 Rev. 023
395).