Fiber View

If, in the Setup page, the parameter Table View is set to Fiber, a screen
as the following one displays:
Figure 129 Result Fiber View
The Fiber View allows to display results of one fiber:
Loss B -> A and loss A -> B at each wavelength
the average loss at each wavelength
the ORL A and ORL B at each wavelength
the distance is displayed on top of the screen

Fault Finder

If, in the Setup page, the Fault Finder parameter is set to Yes, a Fault
analysis is automatically launched, if one value exceeds the thresholds
defined in the Setup menu.
As soon as the Fault Finder finishes its analysis, an easy to interpret table
is displayed. It indicates the attenuation values and/or reflectance of the