Chapter 16 Fiber Characterization
Referencing the test functions
Activate the AP/CD function on the T-BERD/MTS
8000 V2
1
Press the hard key H
your finger or stylus.
Wait for the tuning process of the DISPAP module (~1min.)
2
Press the hard key R
3
Press the soft key

Activate the AP/CD mode of the OBS5x0

1
Press the Application hard key until the AP/CD LED is lighted.
2
Press Output hard key to activate the BBS .
Perform the AP / CD referencing on the T-BERD/MTS
8000 V2
1
Touch the AP / CD tab and press the hard key.
2
Press the soft key Acq. Ref. to enter the reference menu.
3
Select Make Reference to Yes.
4
Press Yes to confirm making a new reference.
5
Drop down using the navigation keys, or touch the BBS Serial
Number parameter, and enter the OBS5x0 serial number using
numerical keyboard and press Enter.
486
and light the icon
OME
ESULTS
until OUT B is selected
User Manual
and touch the MTAU tab
78000010220 Rev. 023
/