YOKOGAWA WT1800 User Manual page 90

Precision power analyzer communication interface
Hide thumbs Also See for WT1800:
Table of Contents

Advertisement

5.13 INPut Group
[:INPut]:CURRent:CONFig:ELEMent<x>
Function
Sets or queries the valid electric current range of
the specified element.
[:INPut]:CURRent:CONFig:ELEMent<x>
Syntax
{ALL|<Current>[,<Current>][,<Current
>]...}
[:INPut]:CURRent:CONFig:ELEMent<x>?
<x> = 1 to 6 (element)
ALL = All ranges are valid.
<Current> = See [:INPut]:CURRent:
CONFig[:ALL]
:INPUT:CURRENT:CONFIG:ELEMENT1 ALL
Example
:INPUT:CURRENT:CONFIG:ELEMENT1? ->
:INPUT:CURRENT:CONFIG:ELEMENT1 ALL
:INPUT:CURRENT:CONFIG:
ELEMENT1 50,10,5,1
:INPUT:CURRENT:CONFIG:ELEMENT1? ->
:INPUT:CURRENT:CONFIG:ELEMENT1 50.0E
+00,10.0E+00,5.0E+00,1.0E+00
Description In the parameters, list all the electric current
ranges that you want to enable. To enable all the
ranges, specify the parameter "ALL."
[:INPut]:CURRent:EXTSensor?
Function
Queries all external current sensor range settings.
[:INPut]:CURRent:EXTSensor?
Syntax
Description This command is only valid on models with the
external current sensor input (/EX) option.
[:INPut]:CURRent:EXTSensor:CONFig?
Function
Queries the valid external current sensor ranges
of all elements.
[:INPut]:CURRent:EXTSensor:CONFig?
Syntax
[:INPut]:CURRent:EXTSensor:
CONFig[:ALL]
Function
Collectively sets the valid external current sensor
range of all elements.
[:INPut]:CURRent:EXTSensor:
Syntax
CONFig[:ALL] {ALL|<Voltage>[,
<Voltage>][,<Voltage>]...}
ALL = All ranges are valid.
• When the crest factor is set to 3:
<Voltage> = 50 mV, 100 mV, 200 mV, 500 mV,
1 V, 2 V, 5 V, 10 V
• When the crest factor is set to 6:
<Voltage> = 25 mV, 50 mV, 100 mV, 250 mV,
500 mV, 1 V, 2.5 V, 5 V
:INPUT:CURRENT:EXTSENSOR:CONFIG:
Example
ALL ALL
:INPUT:CURRENT:EXTSENSOR:CONFIG:
ALL 10,5,2,1
Description In the parameters, list all the external current
sensor ranges that you want to enable. To enable
all the ranges, specify the parameter "ALL."
5-58
[:INPut]:CURRent:EXTSensor:CONFig:
ELEMent<x>
Function
Sets or queries the valid external current sensor
ranges of the specified element.
[:INPut]:CURRent:EXTSensor:CONFig:
Syntax
ELEMent<x> {ALL|<Voltage>[,
<Voltage>][,<Voltage>]...}
[:INPut]:CURRent:EXTSensor:CONFig:
ELEMent<x>?
<x> = 1 to 6 (element)
ALL = All ranges are valid.
<Voltage> = See [:INPut]:CURRent:
EXTSensor:CONFig[:ALL]
:INPUT:CURRENT:EXTSENSOR:CONFIG:
Example
ELEMENT1 ALL
:INPUT:CURRENT:EXTSENSOR:CONFIG:
ELEMENT1? ->
:INPUT:CURRENT:EXTSENSOR:CONFIG:
ELEMENT1 ALL
:INPUT:CURRENT:EXTSENSOR:CONFIG:
ELEMENT1 10,5,2,1
:INPUT:CURRENT:EXTSENSOR:CONFIG:
ELEMENT1? ->
:INPUT:CURRENT:EXTSENSOR:CONFIG:
ELEMENT1 10.00E+00,5.00E+00,
2.00E+00,1.00E+00
Description In the parameters, list all the external current
sensor ranges that you want to enable. To enable
all the ranges, specify the parameter "ALL."
[:INPut]:CURRent:EXTSensor:DISPlay
Function
Sets or queries the display mode of the external
current sensor range.
[:INPut]:CURRent:EXTSensor:DISPlay
Syntax
{DIRect|MEASure}
[:INPut]:CURRent:EXTSensor:DISPlay?
:INPUT:CURRENT:EXTSENSOR:
Example
DISPLAY DIRECT
:INPUT:CURRENT:EXTSENSOR:DISPLAY? ->
:INPUT:CURRENT:EXTSENSOR:
DISPLAY DIRECT
[:INPut]:CURRent:EXTSensor:POJump?
Function
Queries the jump destination ranges of all
elements that are used when a current peak over-
range occurs.
[:INPut]:CURRent:EXTSensor:POJump?
Syntax
IM WT1801-17EN