YOKOGAWA WT1800 User Manual page 109

Precision power analyzer communication interface
Hide thumbs Also See for WT1800:
Table of Contents

Advertisement

:MOTor:SPEed:PRANge
Function
Sets or queries the range of the revolution signal
(pulse input type).
:MOTor:SPEed:PRANge {<NRf>,<NRf>}
Syntax
:MOTor:SPEed:PRANge?
<NRf> = 0.0000 to 99999.9999
:MOTOR:SPEED:PRANGE 10000,0
Example
:MOTOR:SPEED:PRANGE? ->
:MOTOR:SPEED:PRANGE 10000.0000,
0.0000
Description • Set the upper limit and then the lower limit.
• This command is valid when the revolution
signal input type (:MOTor:SPEed:TYPE) is set
to PULSe (pulse input).
:MOTor:SPEed:PULSe
Function
Sets or queries the number of pulses of the
revolution signal (pulse input type).
:MOTor:SPEed:PULSe {<NRf>}
Syntax
:MOTor:SPEed:PULSe?
<NRf> = 1 to 9999
:MOTOR:SPEED:PULSE 60
Example
:MOTOR:SPEED:PULSE? ->
:MOTOR:SPEED:PULSE 60
Description This command is valid when the revolution signal
input type (:MOTor:SPEed:TYPE) is set to PULSe
(pulse input).
:MOTor:SPEed:RANGe
Function
Sets or queries the voltage range of the revolution
signal (analog input type).
:MOTor:SPEed:RANGe {<Voltage>}
Syntax
:MOTor:SPEed:RANGe?
<Voltage> = 1 V, 2 V, 5 V, 10 V,
20 V
:MOTOR:SPEED:RANGE 20V
Example
:MOTOR:SPEED:RANGE? ->
:MOTOR:SPEED:RANGE 20.0E+00
Description This command is valid when the revolution
signal input type (:MOTor:SPEed:TYPE) is set to
ANALog (analog input).
:MOTor:SPEed:SCALing
Function
Sets or queries the rotating speed computation
scaling factor.
:MOTor:SPEed:SCALing {<NRf>}
Syntax
:MOTor:SPEed:SCALing?
<NRf> = 0.0001 to 99999.9999
:MOTOR:SPEED:SCALING 1
Example
:MOTOR:SPEED:SCALING? ->
:MOTOR:SPEED:SCALING 1.0000
IM WT1801-17EN
:MOTor:SPEed:TYPE
Function
Sets or queries the revolution signal input type.
:MOTor:SPEed:TYPE {ANALog|PULSe}
Syntax
:MOTor:SPEed:TYPE?
:MOTOR:SPEED:TYPE ANALOG
Example
:MOTOR:SPEED:TYPE? ->
:MOTOR:SPEED:TYPE ANALOG
:MOTor:SPEed:UNIT
Function
Sets or queries the unit that is added to the
rotating speed computation result.
:MOTor:SPEed:UNIT {<String>}
Syntax
:MOTor:SPEed:UNIT?
<String> = Up to 8 characters
:MOTOR:SPEED:UNIT "rpm"
Example
:MOTOR:SPEED:UNIT? ->
:MOTOR:SPEED:UNIT "rpm"
Description This command has no effect on the computation
result.
:MOTor:SSPeed
Function
Sets or queries the frequency measurement
source for the synchronous speed (SyncSp)
computation.
:MOTor:SSPeed {U<x>|I<x>}
Syntax
:MOTor:SSPeed?
<x> = 1 to 6 (element)
:MOTOR:SSPEED I1
Example
:MOTOR:SSPEED? ->
:MOTOR:SSPEED I1
:MOTor:SYNChronize
Function
Sets or queries the synchronization source for
the rotating speed (Speed) and torque (Torque)
computations.
:MOTor:SYNChronize {U<x>|I<x>|
Syntax
EXTernal|NONE}
:MOTor:SYNChronize?
<x> = 1 to 6 (element)
EXTernal = External clock input (Ext
Clk)
NONE = No synchronization source
:MOTOR:SYNCHRONIZE NONE
Example
:MOTOR:SYNCHRONIZE? ->
:MOTOR:SYNCHRONIZE NONE
:MOTor:TORQue?
Function
Queries all torque (Torque) settings.
:MOTor:TORQue?
Syntax
5.16 MOTor Group
1
2
3
4
5
6
App
Index
5-77

Advertisement

Table of Contents
loading

Table of Contents