5.13 INPut Group
[:INPut]:CURRent:RANGe[:ALL]
Function
Collectively sets the electric current range of all
elements.
[:INPut]:CURRent:RANGe[:ALL] {<Curre
Syntax
nt>|(EXTernal,<Voltage>)}
• 50 A input elements
• When the crest factor is set to 3:
<Current> = 1 A, 2 A, 5 A, 10 A, 20 A, 50 A (for
direct current input)
<Voltage> = 50 mV, 100 mV, 200 mV, 500 mV,
1 V, 2 V, 5 V, 10 V (for external current sensor
input)
• When the crest factor is set to 6:
<Current> = 500 mA, 1 A, 2.5 A, 5 A, 10 A, 25
A (for direct current input)
<Voltage> = 25 mV, 50 mV, 100 mV, 250 mV,
500 mV, 1 V, 2.5 V, 5 V (for external current
sensor input)
• 5 A input elements
• When the crest factor is set to 3:
<Current> = 10 mA, 20 mA, 50 mA, 100 mA,
200 mA, 500 mA, 1 A, 2 A, 5 A (for direct
current input)
<Voltage> = 50 mV, 100 mV, 200 mV, 500 mV,
1 V, 2 V, 5 V, 10 V (for external current sensor
input)
• When the crest factor is set to 6:
<Current> = 5 mA, 10 mA, 25 mA, 50 mA,
100 mA, 250 mA, 500 mA, 1 A, 2.5 A (for direct
current input)
<Voltage> = 25 mV, 50 mV, 100 mV, 250 mV,
500 mV, 1 V, 2.5 V, 5 V (for external current
sensor input)
:INPUT:CURRENT:RANGE:ALL 50A
Example
:INPUT:CURRENT:RANGE:ALL
EXTERNAL,10V
Description EXTernal and <Voltage> can only be selected
on models with the external current sensor input
(/EX) option.
[:INPut]:CURRent:RANGe:ELEMent<x>
Function
Sets or queries the electric current range of the
specified element.
[:INPut]:CURRent:RANGe:
Syntax
ELEMent<x> {<Current>|(EXTernal,
<Voltage>)}
[:INPut]:CURRent:RANGe:ELEMent<x>?
<x> = 1 to 6 (element)
<Current>,<Voltage> = See [:INPut]:
CURRent:RANGe[:ALL]
:INPUT:CURRENT:RANGE:ELEMENT1 50A
Example
:INPUT:CURRENT:RANGE:ELEMENT1? ->
:INPUT:CURRENT:RANGE:
ELEMENT1 50.0E+00
:INPUT:CURRENT:RANGE:
ELEMENT1 EXTERNAL,10V
:INPUT:CURRENT:RANGE:ELEMENT1? ->
:INPUT:CURRENT:RANGE:
ELEMENT1 EXTERNAL,10.00E+00
5-60
[:INPut]:CURRent:RANGe:
{SIGMA|SIGMB|SIGMC}
Function
Collectively sets the electric current range of all
the elements that belong to the specified wiring
unit (ΣA, ΣB, or ΣC).
[:INPut]:CURRent:RANGe:{SIGMA|SIGMB|
Syntax
SIGMC} {<Current>|(EXTernal,
<Voltage>)}
<Current>,<Voltage> = See [:INPut]:
CURRent:RANGe[:ALL]
:INPUT:CURRENT:RANGE:SIGMA 50A
Example
:INPUT:CURRENT:RANGE:
SIGMA EXTERNAL,10V
Description SIGMA, SIGMB, or SIGMC is invalid if the wiring
system setting ([:INPut]:WIRing) is made in such
a way that the corresponding wiring unit (ΣA, ΣB,
or ΣC) does not exist.
[:INPut]:CURRent:SRATio?
Function
Queries the external current sensor conversion
ratios of all elements.
[:INPut]:CURRent:SRATio?
Syntax
Description This command is only valid on models with the
external current sensor input (/EX) option.
[:INPut]:CURRent:SRATio[:ALL]
Function
Collectively sets the external current sensor
conversion ratios of all elements.
[:INPut]:CURRent:SRATio[:ALL]
Syntax
{<NRf>}
<NRf> = 0.0001 to 99999.9999
:INPUT:CURRENT:SRATIO:ALL 10
Example
[:INPut]:CURRent:SRATio:ELEMent<x>
Function
Sets or queries the external current sensor
conversion ratio of the specified element.
[:INPut]:CURRent:SRATio:
Syntax
ELEMent<x> {<NRf>}
[:INPut]:CURRent:SRATio:
ELEMent<x>?
<x> = 1 to 6 (element)
<NRf> = 0.0001 to 99999.9999
:INPUT:CURRENT:SRATIO:ELEMENT1 10
Example
:INPUT:CURRENT:SRATIO:ELEMENT1? ->
:INPUT:CURRENT:SRATIO:
ELEMENT1 10.0000