YOKOGAWA WT1800 User Manual page 95

Precision power analyzer communication interface
Hide thumbs Also See for WT1800:
Table of Contents

Advertisement

[:INPut]:NULL:TARGet:{U<x>|I<x>}
Function
Sets or queries the target of the voltage or current
NULL operation of the specified element.
[:INPut]:NULL:TARGet:{U<x>|
Syntax
I<x>} {ON|HOLD|OFF}
[:INPut]:NULL:TARGet:{U<x>|I<x>}?
<x> = 1 to 6 (element)
ON = NULL feature enabled (When NULL
is set to ON, a new NULL value is
acquired.)
HOLD = NULL feature enabled (When
NULL is set to ON, the previous NULL
value is maintained.)
OFF = NULL feature disabled (NULL
operation is not performed.)
:INPUT:NULL:TARGET:U1 ON
Example
:INPUT:NULL:TARGET:U1? ->
:INPUT:NULL:TARGET:U1 ON
[:INPut]:NULL:TARGet:{UALL|IALL}
Function
Collectively sets the target of the voltage or
current NULL operation of all elements.
[:INPut]:NULL:TARGet:{UALL|
Syntax
IALL} {ON|HOLD|OFF}
ON = NULL feature enabled (When NULL
is set to ON, a new NULL value is
acquired.)
HOLD = NULL feature enabled (When
NULL is set to ON, the previous NULL
value is maintained.)
OFF = NULL feature disabled (NULL
operation is not performed.)
:INPUT:NULL:TARGET:UALL ON
Example
[:INPut]:POVer?
Function
Queries the peak over-range information.
[:INPut]:POVer?
Syntax
:INPUT:POVER? -> 0
Example
Description • The peak over-range information of each
element is mapped as shown below. For the
response, the sum of the values of each bit is
returned in decimal format.
• For example, a response of 16 indicates that a
peak over-range is occurring at U3.
15 14 13 12 11 10 9 8 7 6 5 4 3 2 1 0
A2 A 1 Tq Sp I6 U6 I 5 U5 I 4 U4 I 3 U3 I 2 U2 I1 U1
Sp: Rotating speed
Tq: Torque
A1: AUX1
A2: AUX2
[:INPut]:SCALing?
Function
Queries all scaling settings.
[:INPut]:SCALing?
Syntax
IM WT1801-17EN
[:INPut]:SCALing:STATe?
Function
Queries the on/off statuses of the scaling of all
elements.
[:INPut]:SCALing:STATe?
Syntax
[:INPut]:SCALing[:STATe][:ALL]
Function
Collectively sets the on/off status of the scaling of
all elements.
[:INPut]:SCALing[:STATe][:ALL]
Syntax
{<Boolean>}
:INPUT:SCALING:STATE:ALL OFF
Example
[:INPut]:SCALing[:STATe]:ELEMent<x>
Function
Sets or queries the on/off status of the scaling of
the specified element.
[:INPut]:SCALing[:STATe]:ELEMent<x>
Syntax
{<Boolean>}
[:INPut]:SCALing[:STATe]:ELEMent<x>?
<x> = 1 to 6 (element)
:INPUT:SCALING:STATE:ELEMENT1 OFF
Example
:INPUT:SCALING:STATE:ELEMENT1? ->
:INPUT:SCALING:STATE:ELEMENT1 0
[:INPut]:SCALing:{VT|CT|SFACtor}?
Function
Queries the VT ratios, CT ratios, or power
coefficients of all elements.
[:INPut]:SCALing:{VT|CT|SFACtor}?
Syntax
[:INPut]:SCALing:{VT|CT|SFACtor}
[:ALL]
Function
Collectively sets the VT ratio, CT ratio, or power
coefficient of all elements.
[:INPut]:SCALing:{VT|CT|SFACtor}
Syntax
[:ALL] {<NRf>}
<NRf> = 0.0001 to 99999.9999
:INPUT:SCALING:VT:ALL 1
Example
[:INPut]:SCALing:{VT|CT|SFACtor}:
ELEMent<x>
Function
Sets or queries the VT ratio, CT ratio, or power
coefficient of the specified element.
[:INPut]:SCALing:{VT|CT|SFACtor}:
Syntax
ELEMent<x> {<NRf>}
[:INPut]:SCALing:{VT|CT|SFACtor}:
ELEMent<x>?
<x> = 1 to 6 (element)
<NRf> = 0.0001 to 99999.9999
:INPUT:SCALING:VT:ELEMENT1 1
Example
:INPUT:SCALING:VT:ELEMENT1? ->
:INPUT:SCALING:VT:ELEMENT1 1.0000
5.13 INPut Group
1
2
3
4
5
6
App
Index
5-63