YOKOGAWA WT1800 User Manual page 110

Precision power analyzer communication interface
Hide thumbs Also See for WT1800:
Table of Contents

Advertisement

5.16 MOTor Group
:MOTor:TORQue:AUTO
Function
Sets or queries the voltage auto range on/off
status of the torque signal (analog input type).
:MOTor:TORQue:AUTO {<Boolean>}
Syntax
:MOTor:TORQue:AUTO?
:MOTOR:TORQUE:AUTO ON
Example
:MOTOR:TORQUE:AUTO? ->
:MOTOR:TORQUE:AUTO 1
Description This command is valid when the torque signal
input type (:MOTor:TORQue:TYPE) is set to
ANALog (analog input).
:MOTor:TORQue:LSCale?
Function
Queries all torque signal (analog input type) linear
scaling settings.
:MOTor:TORQue:LSCale?
Syntax
Description Linear scaling settings are valid when the torque
signal input type (:MOTor:TORQue:TYPE) is set
to ANALog (analog input).
:MOTor:TORQue:LSCale:AVALue
Function
Sets or queries the slope (A) of the linear scale of
the torque signal (analog input type).
:MOTor:TORQue:LSCale:AVALue {<NRf>}
Syntax
:MOTor:TORQue:LSCale:AVALue?
<NRf> = 1.000E-03 to 1.000E+06
:MOTOR:TORQUE:LSCALE:AVALUE 1.000
Example
:MOTOR:TORQUE:LSCALE:AVALUE? ->
:MOTOR:TORQUE:LSCALE:
AVALUE 1.000E+00
:MOTor:TORQue:LSCale:BVALue
Function
Sets or queries the offset (B) of the linear scale of
the torque signal (analog input type).
:MOTor:TORQue:LSCale:BVALue {<NRf>}
Syntax
:MOTor:TORQue:LSCale:BVALue?
<NRf> = -1.000E+06 to 1.000E+06
:MOTOR:TORQUE:LSCALE:BVALUE 0
Example
:MOTOR:TORQUE:LSCALE:BVALUE? ->
:MOTOR:TORQUE:LSCALE:
BVALUE 0.000E+00
:MOTor:TORQue:LSCale:CALCulate?
Function
Queries all parameter calculation settings for the
linear scale of the torque signal (analog input
type).
:MOTor:TORQue:LSCale:CALCulate?
Syntax
5-78
:MOTor:TORQue:LSCale:CALCulate:
{P1X|P1Y|P2X|P2Y}
Function
Sets or queries the data (Point1X, Point1Y,
Point2X, or Point2Y) for parameter calculations of
the linear scale of the torque signal (analog input
type).
:MOTor:TORQue:LSCale:CALCulate:
Syntax
{P1X|P1Y|P2X|P2Y} {<NRf>}
:MOTor:TORQue:LSCale:CALCulate:
{P1X|P1Y|P2X|P2Y}?
<NRf> = -1.000E+12 to 1.000E+12
:MOTOR:TORQUE:LSCALE:CALCULATE:P1X 0
Example
:MOTOR:TORQUE:LSCALE:CALCULATE:
P1X? ->
:MOTOR:TORQUE:LSCALE:CALCULATE:
P1X 0.000E+00
:MOTor:TORQue:LSCale:CALCulate:
EXECute
Function
Calculates parameters for the linear scale of the
torque signal (analog input type).
:MOTor:TORQue:LSCale:CALCulate:
Syntax
EXECute
:MOTOR:TORQUE:LSCALE:CALCULATE:
Example
EXECUTE
Description The WT1800 uses the data that has been
specified (Point1X, Point1Y, Point2X, and
Point2Y) to calculate and set the slope (A) and
offset (B) of the linear scale.
:MOTor:TORQue:PRANge
Function
Sets or queries the range of the torque signal
(pulse input type).
:MOTor:TORQue:PRANge {<NRf>,<NRf>}
Syntax
:MOTor:TORQue:PRANge?
<NRf> = -10000.0000 to 10000.0000
:MOTOR:TORQUE:PRANGE 50,-50
Example
:MOTOR:TORQUE:PRANGE? ->
:MOTOR:TORQUE:PRANGE 50.0000,
-50.0000
Description • Set the upper limit and then the lower limit.
• This command is valid when the torque signal
input type (:MOTor:TORQue:TYPE) is set to
PULSe (pulse input).
:MOTor:TORQue:RANGe
Function
Sets or queries the voltage range of the torque
signal (analog input type).
:MOTor:TORQue:RANGe {<Voltage>}
Syntax
:MOTor:TORQue:RANGe?
<Voltage> = 1 V, 2 V, 5 V, 10 V,
20 V
:MOTOR:TORQUE:RANGE 20V
Example
:MOTOR:TORQUE:RANGE? ->
:MOTOR:TORQUE:RANGE 20.0E+00
Description This command is valid when the torque signal
input type (:MOTor:TORQue:TYPE) is set to
ANALog (analog input).
IM WT1801-17EN

Advertisement

Table of Contents
loading

Table of Contents