Communication Query Errors (400 to 499)
Code
Message
410
Query INTERRUPTED.
420
Query UNTERMINATED.
430
Query DEADLOCKED.
440
Query UNTERMINATED after indefinite response. Do not write a query after *IDN? or *OPT?.
System Communication Errors (300 and 399)
Code
Message
300
Communication device-specific error.
399
Fatal error in the communication driver.
Communication Warning (50)
Code
Message
50
*OPC/? exists in message.
Other Error (350)
Code
Message
350
Queue overflow.
Note
Code 350 occurs when the error queue overflows. This error is only returned in response to a :STATus:
ERRor? query; it is never displayed on the screen.
IM WT1801-17EN
Corrective Action
Check the transmission and reception order.
Check the transmission and reception order.
Keep program messages (including <PMT>) to 1024
bytes or less in length.
Corrective Action
Servicing required.
Servicing required.
Corrective Action
Write *OPC or *OPC? at the end of program messages. —
Corrective Action
Read the error queue.