YOKOGAWA WT1800 User Manual page 85

Precision power analyzer communication interface
Hide thumbs Also See for WT1800:
Table of Contents

Advertisement

:HSPeed:RECord:ITEM:{U<x>|UA|UB|UC}
Function
Sets or queries whether the specified element or
wiring unit (ΣA, ΣB, or ΣC) of the numeric data
(voltage) will be saved.
:HSPeed:RECord:ITEM:{U<x>|UA|UB|
Syntax
UC} {<Boolean>}
:HSPeed:RECord:ITEM:{U<x>|UA|UB|UC}?
<x> = 1 to 6 (element)
UA, UB, UC = UΣA, UΣB, UΣC
:HSPEED:RECORD:ITEM:U1 ON
Example
:HSPEED:RECORD:ITEM:U1? ->
:HSPEED:RECORD:ITEM:U1 1
:HSPeed:RECord:ITEM:PRESet:ALL
Function
Sets, at the same time, whether all numeric data
items will be saved.
:HSPeed:RECord:ITEM:PRESet:
Syntax
ALL {<Boolean>}
:HSPEED:RECORD:ITEM:PRESET:ALL ON
Example
:HSPeed:RECord:ITEM:PRESet:{ELEMent<
x>|SIGMA|SIGMB|SIGMC}
Function
Sets, at the same time, whether the specified
element or wiring unit (ΣA, ΣB, or ΣC) of all the
types of numeric data will be saved.
:HSPeed:RECord:ITEM:PRESet:
Syntax
{ELEMent<x>|SIGMA|SIGMB|
SIGMC} {<Boolean>}
<x> = 1 to 6
:HSPEED:RECORD:ITEM:PRESET:
Example
ELEMENT1 ON
:HSPeed:RECord:ITEM:
PRESet:{U|I|P|MOTor|AUX}
Function
Sets, at the same time, whether the specified
functions of all types of numeric data will be saved.
:HSPeed:RECord:ITEM:PRESet:{U|I|P|
Syntax
MOTor|AUX} {<Boolean>}
:HSPEED:RECORD:ITEM:PRESET:U ON
Example
Description • MOTor is only valid on models with the motor
evaluation function (/MTR) option.
• AUX is only valid on models with the auxiliary
input (/AUX) option.
:HSPeed:RECord[:STATe]
Function
Sets or queries whether acquired numeric data is
saved to a file.
:HSPeed:RECord[:STATe] {<Boolean>}
Syntax
:HSPeed:RECord:STATe?
:HSPEED:RECORD:STATE ON
Example
:HSPEED:RECORD:STATE? ->
:HSPEED:RECORD:STATE 1
IM WT1801-17EN
:HSPeed:STARt
Function
Starts data capturing.
:HSPeed:STARt
Syntax
:HSPEED:START
Example
:HSPeed:STATe?
Function
Queries the status of high speed data capturing.
:HSPeed:STATe?
Syntax
::HSPEED:STATE? -> READY
Example
Description • The response is as follows:
INIT = High-speed data capturing is being
initialized.
READy = High-speed data capturing is in stand
by.
STARt = High-speed data capturing is being
performed.
OFF = The WT1800 is not in high speed data
capturing mode.
:HSPeed:STOP
Function
Stops data capturing.
:HSPeed:STOP
Syntax
:HSPEED:STOP
Example
:HSPeed:TRIGger?
Function
Queries all high speed data capturing trigger
settings.
:HSPeed:TRIGger?
Syntax
:HSPeed:TRIGger:LEVel
Function
Sets or queries the trigger level.
:HSPeed:TRIGger:LEVel {<NRf>}
Syntax
:HSPeed:TRIGger:LEVel?
<NRf> = ?100.0 to 100.0(%)
:HSPEED:TRIGGER:LEVEL 0
Example
:HSPEED:TRIGGER:LEVEL? -> :HSPEED:
TRIGGER:LEVEL 0.0
Description This command performs the same setting as the ":
DISPlay:WAVE:TRIGger:LEVel" command.
:HSPeed:TRIGger:MODE
Function
Sets or queries the trigger mode.
:HSPeed:TRIGger:MODE {AUTO|NORMal|
Syntax
OFF}
:HSPeed:TRIGger:MODE?
:HSPEED:TRIGGER:MODE AUTO
Example
:HSPEED:TRIGGER:MODE? ->
:HSPEED:TRIGGER:MODE AUTO
Description This command performs the same setting as the ":
DISPlay:WAVE:TRIGger:MODE" command.
5.11 HSPeed Group
App
Index
5-53
1
2
3
4
5
6

Advertisement

Table of Contents
loading

Table of Contents