Appendix

Appendix 1
This section explains communication error messages.
• On the WT1800, the messages can be displayed in the language that you specify with the Menu
Language setting on the System Config menu. However, any messages that you read from a PC
or other controller will be displayed in English.
• If servicing is necessary to solve the problem indicated by a message, contact your nearest
YOKOGAWA dealer.
• Only communication error messages are listed here. For details on other error messages, see
the user's manual, IM WT1801-02EN.
• Communication syntax errors
• Communication execution errors
• Device-specific and other errors
• Communication query errors
• System communication errors
Communication Syntax Errors (100 to 199)
Code
Message
102
Syntax error.
103
Invalid separator.
104
Data type error.
108
Parameter not allowed.
109
Missing parameter.
111
Header separator error.
112
Program mnemonic too long.
113
Undefined header.
114
Header suffix out of range.
120
Numeric data error.
123
Exponent too large.
124
Too many digits.
128
Numeric data not allowed.
131
Invalid suffix.
134
Suffix too long.
138
Suffix not allowed.
141
Invalid character data.
144
Character data too long.
148
Character data not allowed.
150
String data error.
151
Invalid string data.
IM WT1801-17EN

Error Messages

100 to 199
200 to 299
300 to 399
400 to 499
300 and 399
Corrective Action
A syntax error not covered by error codes 100 to 199.
Separate data values with a comma.
See page 4-6 and 4-7 and use the correct data type for
each parameter.
Check the number of data values.
Be sure to include all necessary data values.
Use a comma to separate each header from its data.
Check the length of the mnemonic (alphanumeric
character string).
Check the header.
Check the header.
A value must be specified where the syntax contains
<NRf>.
Where the syntax contains <NR3>, make the exponent
that follows E smaller.
Limit numeric values to 255 digits or less.
Use a data type other than <NRf>.
Check the units where the syntax contains <Voltage>,
<Current>, <Time>, or <Frequency>.
Check the units where the syntax contains <Voltage>,
<Current>, <Time>, or <Frequency>.
Units of measurement can only be used where the
syntax contains <Voltage>, <Current>, <Time>, or
<Frequency>.
Be sure to select one of the listed choices when the
syntax contains {...|...|...}.
Check the spelling of the strings when the syntax
contains {...|...|...}.
Use a data type other than {...|...|...}.
Enclose parameters with single or double quotation
marks where the syntax contains <string data>.
The parameter is either too long, or it contains an
unusable character.
Listed below
1
2
3
4
5
6
Page
Chapters 4
and
5
App
4-1
4-6 and
4-7
4-6 and
Index
chapter 5
4-6 and
chapter 5
4-1
Chapter 5
Chapter 5
Chapter 5
4-6
4-6 and
chapter 5
4-6 and
chapter 5
4-6 and
chapter 5
4-6
4-6
4-6
Chapters 4
and
5
Chapter 5
Chapter 5
4-7
Chapter 5
App-1