YOKOGAWA WT1800 User Manual page 128

Precision power analyzer communication interface
Hide thumbs Also See for WT1800:
Table of Contents

Advertisement

5.20 STORe Group
:STORe:NUMeric[:NORMal]:PRESet<x>
Function
Presets the output on/off pattern of the element
functions to be used when numeric data is stored.
:STORe:NUMeric[:NORMal]:PRESet<x>
Syntax
<x> = 1 or 2 (preset number)
:STORE:NUMERIC:NORMAL:PRESET1
Example
Description For details on the storage item setting patterns
that result when the pattern is reset, see the
features guide, IM WT1801-01EN.
:STORe:RESet
Function
Resets the numeric data storage feature.
:STORe:RESet
Syntax
:STORE:RESET
Example
:STORe:RTIMe?
Function
Queries the storage start and end times for real-
time storage mode.
:STORe:RTIMe?
Syntax
:STORe:RTIMe:{STARt|END}
Function
Sets or queries the storage start or end time for
real-time storage mode.
:STORe:RTIMe:{STARt|END} {<NRf>,
Syntax
<NRf>,<NRf>,<NRf>,<NRf>,<NRf>}
:STORe:RTIMe:{STARt|END}?
{<NRf>,<NRf>,<NRf>,<NRf>,
<NRf>,<NRf>} = 2001,1,1,0,0,0 to
2099,12,31,23,59,59
First <NRf> = 2001 to 2099 (year)
Second <NRf> = 1 to 12 (month)
Third <NRf> = 1 to 31 (day)
Fourth <NRf> = 0 to 23 (hour)
Fifth <NRf> = 0 to 59 (minute)
Sixth <NRf> = 0 to 59 (second)
:STORE:RTIME:START 2011,1,1,0,0,0
Example
:STORE:RTIME:START? ->
:STORE:RTIME:START 2011,1,1,0,0,0
Description This command is valid when the storage mode
(:STORe:SMODe) is set to RTIMe.
:STORe:SASTart
Function
Sets or queries whether numeric data is stored
when storage starts.
:STORe:SASTart {<Boolean>}
Syntax
:STORe:SASTart?
:STORE:SASTART OFF
Example
:STORE:SASTART? ->
:STORE:SASTART 0
Description This command is valid when the storage mode
(:STORe:SMODe) is set to MANual, RTIMe, or
INTEGrate (and when the storage interval is a
value other than 0 for MANual or RTIMe mode).
5-96
:STORe:SMODe
Function
Sets or queries the storage mode.
:STORe:SMODe {MANual|RTIMe|
Syntax
INTEGrate|EVENt|SINGle}
:STORe:SMODe?
MANual = Manual storage mode
RTIMe = Real-time storage mode
INTEGrate = Integration-synchronized
storage mode
EVENt = Event-synchronized storage
mode
SINGle = Single-shot storage mode
:STORE:SMODE MANUAL
Example
:STORE:SMODE? ->
:STORE:SMODE MANUAL
:STORe:STARt
Function
Begins the storing of numeric data.
:STORe:STARt
Syntax
:STORE:START
Example
Description If :STORe:SMODe is set to MANual, the storage
operation is executed.If :STORe:SMODe is set to
RTIMe, INTEGrate, or EVENt, the WT1800 enters
into a storage wait state.If :STORe:SMODe is set
to SINGle, the storage operation is executed.
:STORe:STATe?
Function
Sets or queries the storage state.
:STORe:STATe?
Syntax
:STORE:STATE? -> RESET
Example
Description • The response is as follows:
RESet = Storage reset
READy = Storage standby
STARt = Currently storing
STOP = Storage stopped
COMPlete = Storage complete (or
ended due to an error)
CONVert = Converting stored data to
CSV format
CLOSe = Final processing of a WTS or
HDS file
:STORe:STOP
Function
Stops the storing of numeric data.
:STORe:STOP
Syntax
:STORE:STOP
Example
:STORe:TEVent
Function
Sets or queries the event that the event-
synchronized storage mode will trigger on.
:STORe:TEVent {<NRf>}
Syntax
:STORe:TEVent?
<NRf> = 1 to 8 (event number)
:STORE:TEVENT 1
Example
:STORE:TEVENT? ->
:STORE:TEVENT 1
Description This command is valid when the storage mode
(:STORe:SMODe) is set to EVENt.
IM WT1801-17EN

Advertisement

Table of Contents
loading

Table of Contents