Aoutput Group; Front Panel - YOKOGAWA WT1800 User Manual

Precision power analyzer communication interface
Hide thumbs Also See for WT1800:
Table of Contents

Advertisement

5.2

AOUTput Group

The commands in this group deal with D/A output.
You can make the same settings and queries that you can make by pressing UTILITY on the front panel, and then
using the D/A Output Items menu.
The commands in this group are only valid on models with the D/A output (/DA) option.
:AOUTput?
Function
Queries all D/A output settings.
:AOUTput?
Syntax
:AOUTput:NORMal?
Function
Queries all D/A output settings.
:AOUTput:NORMal?
Syntax
:AOUTput[:NORMal]:CHANnel<x>
Function
Sets or queries a D/A output item (function,
element, or harmonic order).
:AOUTput[:NORMal]:CHANnel<x> {NONE|
Syntax
<Function>[,<Element>][,<Order>]}
:AOUTput[:NORMal]:CHANnel<x>?
<x> = 1 to 20 (output channel)
NONE = No output item
<Function> = {URMS|IRMS|P|S|Q|...}
<Element> = {<NRf>|SIGMa|SIGMB|SIGMC}
(<NRf> = 1 to 6)
<Order> = {TOTal|DC|<NRf>}
(<NRf> = 1 to 500)
:AOUTPUT:NORMAL:CHANNEL1 URMS,1
Example
:AOUTPUT:NORMAL:CHANNEL1? ->
:AOUTPUT:NORMAL:CHANNEL1 URMS,1
:AOUTPUT:NORMAL:CHANNEL1 UK,1,1
:AOUTPUT:NORMAL:CHANNEL1? ->
:AOUTPUT:NORMAL:CHANNEL1 UK,1,1
Description • For information about the options available for
<Function>, see Function Option List (1), in the
DISPlay Group section on page 5-38.
• If <Element> is omitted, the element is set to 1.
• If <Order> is omitted, the order is set to TOTal.
• <Element> and <Order> are omitted from
responses to functions that do not need them.
5-16
:AOUTput[:NORMal]:IRTime
Function
Sets or queries the integration time that is used in
the D/A output of the integrated value.
:AOUTput[:NORMal]:IRTime
Syntax
{<NRf>,<NRf>,<NRf>}
:AOUTput[:NORMal]:IRTime?
{<NRf>,<NRf>,<NRf>} = 0,0,0 to
10000,0,0
First <NRf> = 0 to 10000 (hours)
Second <NRf> = 0 to 59 (minutes)
Third <NRf> = 0 to 59 (seconds)
:AOUTPUT:NORMAL:IRTIME 1,0,0
Example
:AOUTPUT:NORMAL:IRTIME? ->
:AOUTPUT:NORMAL:IRTIME 1,0,0
:AOUTput[:NORMal]:MODE<x>
Function
Sets or queries the rated-value setup mode for
D/A output items.
:AOUTput[:NORMal]:
Syntax
MODE<x>{FIXed|MANual}
:AOUTput[:NORMal]:MODE<x>?
<x> = 1 to 20 (output channel)
:AOUTPUT:NORMAL:MODE1 FIXED
Example
:AOUTPUT:NORMAL:MODE1? ->
:AOUTPUT:NORMAL:MODE1 FIXED
:AOUTput[:NORMal]:RATE<x>
Function
Sets or queries the rated maximum or minimum
value for D/A output items.
:AOUTput[:NORMal]:RATE<x>{<NRf>,
Syntax
<NRf>}
:AOUTput[:NORMal]:RATE<x>?
<x> = 1 to 20 (output channel)
<NRf> = –9.999E+12 to 9.999E+12
:AOUTPUT:NORMAL:RATE1 100,-100
Example
:AOUTPUT:NORMAL:RATE1? ->
:AOUTPUT:NORMAL:
RATE1 100.0E+00,-100.0E+00
Description • Set the upper limit and then the lower limit.
• This setting is valid when the D/A output rated-
value setup mode (:AOUTput [:NORMal]:
MODE<x>) is set to MANual.
IM WT1801-17EN

Advertisement

Table of Contents
loading

Table of Contents